Initial performance evaluation of a high resolution Albira small animal positron emission tomography scanner with monolithic crystals and depth-of-interaction encoding from a user&apos, s perspective

  • Balcerzyk, Marcin
  • Kontaxakis, George
  • Delgado, Mercedes
  • Garcia-Garcia, Luis
  • Correcher Salvador, Carlos
  • Gonzalez, Aurora
  • Rubio, Jose L.
  • Pozo, Miguel A.
  • González Martínez, Antonio Javier
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
The user evaluation of the Albira small animal positron emission tomography scanner is presented. The system has a circle divide 80 mm x 40 mm field of view. In the center of the field of view, it has 2.49% sensitivity, with 33% solid angle coverage, and 0.88% sensitivity for a mouse phantom. The maximum of noise equivalent counts is 14.6 kcps at 6.0 MBq for a mouse phantom. The scanner employs an innovative crystal design of eight 40(50) x 40(50) x 9.8 mm(3) LYSO tapered monolithic crystals forming detector modules read by position-sensitive photomultipliers. This design allows for easy depth-of-interaction readout. The system saturates at 6.7 MBq for the mouse phantom. As an example of the application of the Albira system, some preliminary results from a study of ischemic stroke on animal models with 18-FDG co-registered with MRI images are shown, The study was supported in part by the MULTIMAG (S-BIO-0170-2006) project of Comunidad de Madrid, Spain, and FPA2007-65013-C02-01 project. MB was supported by Marie-Curie fellowship project no MEIF-CT-2006-041482 from European Commission. JLR is grateful for the support by Fellowship FPI (BES-2005-8126), Ministerio de Educacion y Ciencia de Espana.