Repisalud
Publication
Artículo científico (JournalArticle)
Real-Time 3D PET Image with Pseudoinverse Reconstruction
Real-time positron emission tomography (PET) may provide information from first-shot images, enable PET-guided biopsies, and allow awake animal studies. Fully-3D iterative reconstructions yield the best images in PET, but they are too slow for real-time imaging. Analytical methods such as Fourier back projection (FBP) are very fast, but yield images of poor quality with artifacts due to noise or data incompleteness. In this work, an image reconstruction based on the pseudoinverse of the system response matrix (SRM) is presented. w. To implement the pseudoinverse method, the reconstruction problem is separated into two stages. First, the axial part of the SRM is pseudo-inverted (PINV) to rebin the 3D data into 2D datasets. Then, the resulting 2D slices can be reconstructed with analytical methods or by applying the pseudoinverse algorithm again. The proposed two-step PINV reconstruction yielded good-quality images at a rate of several frames per second, compatible with real time applications. Furthermore, extremely fast direct PINV reconstruction of projections of the 3D image collapsed along specific directions can be implemented.
Project:
ES / FPA2015-65035-P Not available
,
ES / RTI2018-095800-A-I00 Not available
,
ES / S2013 MIT-3024 TOPUS-CM
,
ES / SEV-2015-0505 BMD-3888 PRONTO-CM
,
European Commission / Horizon 2020 Framework Programme Not available
,
European Commission / Horizon 2020 Framework Programme 654002
,
NIHR01CA215700-2