Pasar al contenido principal

Resultados relacionados

Resultados totales (Incluyendo duplicados): 8
Encontrada(s) 1 página(s)

Repositorio Digital de la UPF
Publicación Artículo científico (JournalArticle)

Image formation by linear and nonlinear digital scanned light-sheet fluorescence microscopy with Gaussian and Bessel beam profiles

  • Olarte, Omar E.
  • Licea-Rodriguez, Jacob
  • Palero, Jonathan A.
  • Gualda, Emilio J.
  • Artigas, David
  • Mayer, Jürgen, 1977-
  • Swoger, Jim
  • Sharpe, James
  • Rocha-Mendoza, Israel
  • Rangel-Rojo, Raul
  • Loza-Alvarez, Pablo
We present the implementation of a combined digital scanned light-sheet microscope (DSLM) able to work in the linear and nonlinear regimes under either Gaussian or Bessel beam excitation schemes. A complete characterization of the setup is performed and a comparison of the performance of each DSLM imaging modality is presented using in vivoCaenorhabditis elegans samples. We found that the use of Bessel beam nonlinear excitation results in better image contrast over a wider field of view., This work is supported by the Generalitat de Catalunya grant 2009-SGR-159, the Spanish government grants TEC2009-09698 and SAF2008-00211, the EU project STELUM (FP7-PEOPLE-2007-3-1-IAPP, 217997), the Swiss National Science Foundation (SNSF) Sinergia project grant CRSII3_125447, the FP7-NMP VIBRANT project grant CP-IP 228933-2, the NoE P4L and Laserlab optobio. Jacob Licea-Rodriguez acknowledges CONACYT Mexico for supporting his stay at ICFO through a mixed fellowship. This research has been partially supported by Fundació Cellex Barcelona and was mainly conducted at ICFO’s “Super-Resolution Light Microscopy and Nanoscopy Facility”
UPCommons. Portal del coneixement obert de la UPC
Publicación Artículo científico (JournalArticle)

Third-harmonic generation for the study of Caenorhabditis elegans embryogenesis

  • Aviles Espinosa, Rodrigo
  • Santos, Susana I.C.O.
  • Brodschelm, Andreas
  • Kaenders, Wilhelm G.
  • Alonso Ortega, Cesar
  • Artigas García, David|||0000-0001-5508-2873
  • Loza Álvarez, Pablo
Live microscopy techniques i.e., differential interference contrast, confocal microscopy, etc. have enabled the understanding of the mechanisms involved in cells and tissue formation. In long-term studies, special care must be taken in order to avoid sample damage, restricting the applicability of the different microscopy techniques.We demonstrate the potential of using third-harmonic generation THG microscopy for morphogenesis/embryogenesis studies in living Caenorhabditis elegans C. elegans . Moreover, we show that the THG signal is obtained in all the embryo development stages, showing different tissue/structure information. For this research, we employ a 1550-nm femtosecond fiber laser and demonstrate that the expected water absorption at this wavelength does not severely compromise sample viability. Additionally, this has the important advantage that the THG signal is emitted at visible wavelengths 516 nm . Therefore, standard collection optics and detectors operating near maximum efficiency enable an optimal signal reconstruction. All this, to the best of our knowledge, demonstrates for the first time the noninvasiveness and strong potential of this particular wavelength to be used for highresolution four-dimensional imaging of embryogenesis using unstained C. elegans in vivo samples.
UPCommons. Portal del coneixement obert de la UPC
Publicación Artículo científico (JournalArticle)

Compact ultrafast semiconductor disk laser: targeting GFP based nonlinear applications in living organisms

  • Loza Álvarez, Pablo
  • Artigas García, David|||0000-0001-5508-2873
  • Santos, Susana I.C.O.
  • Aviles Espinosa, Rodrigo
  • Filippidis, George
  • Hamilton, Craig
  • Malcolm, Graeme
  • Weingarten, Kurt J.
  • Südmeyer, Thomas
  • Barbarin, Yohan
  • Keller, Ursula
We present a portable ultrafast Semiconductor Disk Laser (SDL) (or vertical extended cavity surface emitting laser—VECSELs), to be used for nonlinear microscopy. The SDL is modelocked using a quantum-dot semiconductor saturable absorber mirror (SESAM), delivering an average output power of 287 mW, with 1.5 ps pulses at 500 MHz and a central wavelength of 965 nm. Specifically, despite the fact of having long pulses and high repetition rates, we demonstrate the potential of this laser for Two-Photon Excited Fluorescence (TPEF) imaging of in vivo Caenorhabditis elegans (C. elegans) expressing Green Fluorescent Protein (GFP) in a set of neuronal processes and cell bodies. Efficient TPEF imaging is achieved due to the fact that this wavelength matches the peak of the two-photon action cross section of this widely used fluorescent marker. The SDL extended versatility is shown by presenting Second Harmonic Generation images of pharynx, uterus, body wall muscles and its potential to be used to excite other different commercial dyes. Importantly this non-expensive, turn-key, compact laser system could be used as a platform to develop portable nonlinear bio-imaging devices., Peer Reviewed
UPCommons. Portal del coneixement obert de la UPC
Publicación Artículo científico (JournalArticle)

Measurement and correction of in vivo sample aberrations employing a nonlinear guide-star in two-photon excited fluorescence microscopy

  • Aviles Espinosa, Rodrigo
  • Andilla, Jordi
  • Porcar-Guezenec, Rafael
  • Olarte, Omar E.
  • Nieto, Marta
  • Levecq, Xavier
  • Artigas García, David|||0000-0001-5508-2873
  • Loza Álvarez, Pablo
We demonstrate that sample induced aberrations can be measured in a nonlinear microscope. This uses the fact that two-photon excited fluorescence naturally produces a localized point source inside the sample: the nonlinear guide-star (NL-GS). The wavefront emitted from the NL-GS can then be recorded using a Shack-Hartmann sensor. Compensation of the recorded sample aberrations is performed by the deformable mirror in a single-step. This technique is applied to fixed and in vivo biological samples, showing, in some cases, more than one order of magnitude improvement in the total collected signal intensity., Peer Reviewed
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Measurement and correction of in vivo sample aberrations employing a nonlinear guide-star in two-photon excited fluorescence microscopy

  • Aviles Espinosa, Rodrigo
  • Andilla, Jordi
  • Porcar-Guezenec, Rafael
  • Olarte, Omar E.
  • Nieto, Marta
  • Levecq, Xavier
  • Artigas García, David
  • Loza Álvarez, Pablo
We demonstrate that sample induced aberrations can be measured in a nonlinear microscope. This uses the fact that two-photon excited fluorescence naturally produces a localized point source inside the sample: the nonlinear guide-star (NL-GS). The wavefront emitted from the NL-GS can then be recorded using a Shack-Hartmann sensor. Compensation of the recorded sample aberrations is performed by the deformable mirror in a single-step. This technique is applied to fixed and in vivo biological samples, showing, in some cases, more than one order of magnitude improvement in the total collected signal intensity., Peer Reviewed, Postprint (published version)
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Compact ultrafast semiconductor disk laser: targeting GFP based nonlinear applications in living organisms

  • Loza Álvarez, Pablo
  • Artigas García, David
  • Santos, Susana I.C.O.
  • Aviles Espinosa, Rodrigo
  • Filippidis, George
  • Hamilton, Craig
  • Malcolm, Graeme
  • Weingarten, Kurt J.
  • Südmeyer, Thomas
  • Barbarin, Yohan
  • Keller, Ursula
We present a portable ultrafast Semiconductor Disk Laser (SDL) (or vertical extended cavity surface emitting laser—VECSELs), to be used for nonlinear microscopy. The SDL is modelocked using a quantum-dot semiconductor saturable absorber mirror (SESAM), delivering an average output power of 287 mW, with 1.5 ps pulses at 500 MHz and a central wavelength of 965 nm. Specifically, despite the fact of having long pulses and high repetition rates, we demonstrate the potential of this laser for Two-Photon Excited Fluorescence (TPEF) imaging of in vivo Caenorhabditis elegans (C. elegans) expressing Green Fluorescent Protein (GFP) in a set of neuronal processes and cell bodies. Efficient TPEF imaging is achieved due to the fact that this wavelength matches the peak of the two-photon action cross section of this widely used fluorescent marker. The SDL extended versatility is shown by presenting Second Harmonic Generation images of pharynx, uterus, body wall muscles and its potential to be used to excite other different commercial dyes. Importantly this non-expensive, turn-key, compact laser system could be used as a platform to develop portable nonlinear bio-imaging devices., Peer Reviewed, Postprint (published version)
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Image formation by linear and nonlinear digital scanned light-sheet fluorescence microscopy with Gaussian and Bessel beam profiles

  • Olarte, Omar E.
  • Licea-Rodriguez, Jacob
  • Palero, Jonathan A.
  • Gualda, Emilio J.
  • Artigas, David
  • Mayer, Jürgen, 1977-
  • Swoger, Jim
  • Sharpe, James
  • Rocha-Mendoza, Israel
  • Rangel-Rojo, Raul
  • Loza-Alvarez, Pablo
We present the implementation of a combined digital scanned light-sheet microscope (DSLM) able to work in the linear and nonlinear regimes under either Gaussian or Bessel beam excitation schemes. A complete characterization of the setup is performed and a comparison of the performance of each DSLM imaging modality is presented using in vivoCaenorhabditis elegans samples. We found that the use of Bessel beam nonlinear excitation results in better image contrast over a wider field of view., This work is supported by the Generalitat de Catalunya grant 2009-SGR-159, the Spanish government grants TEC2009-09698 and SAF2008-00211, the EU project STELUM (FP7-PEOPLE-2007-3-1-IAPP, 217997), the Swiss National Science Foundation (SNSF) Sinergia project grant CRSII3_125447, the FP7-NMP VIBRANT project grant CP-IP 228933-2, the NoE P4L and Laserlab optobio. Jacob Licea-Rodriguez acknowledges CONACYT Mexico for supporting his stay at ICFO through a mixed fellowship. This research has been partially supported by Fundació Cellex Barcelona and was mainly conducted at ICFO’s “Super-Resolution Light Microscopy and Nanoscopy Facility”
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Third-harmonic generation for the study of Caenorhabditis elegans embryogenesis

  • Aviles Espinosa, Rodrigo
  • Santos, Susana I.C.O.
  • Brodschelm, Andreas
  • Kaenders, Wilhelm G.
  • Alonso Ortega, Cesar
  • Artigas García, David
  • Loza Álvarez, Pablo
Live microscopy techniques i.e., differential interference contrast, confocal microscopy, etc. have enabled the understanding of the mechanisms involved in cells and tissue formation. In long-term studies, special care must be taken in order to avoid sample damage, restricting the applicability of the different microscopy techniques.We demonstrate the potential of using third-harmonic generation THG microscopy for morphogenesis/embryogenesis studies in living Caenorhabditis elegans C. elegans . Moreover, we show that the THG signal is obtained in all the embryo development stages, showing different tissue/structure information. For this research, we employ a 1550-nm femtosecond fiber laser and demonstrate that the expected water absorption at this wavelength does not severely compromise sample viability. Additionally, this has the important advantage that the THG signal is emitted at visible wavelengths 516 nm . Therefore, standard collection optics and detectors operating near maximum efficiency enable an optimal signal reconstruction. All this, to the best of our knowledge, demonstrates for the first time the noninvasiveness and strong potential of this particular wavelength to be used for highresolution four-dimensional imaging of embryogenesis using unstained C. elegans in vivo samples., Postprint (published version)