Endoluminal radiofrequency ablation of the main pancreatic duct is a secure and effective method to produce pancreatic atrophy and to achieve stump closure

  • Andaluz, Anna
  • Ewertowska, Elżbieta
  • Moll, Xavier
  • Aguilar, Adrià
  • García Arnas, Félix
  • Fondevila, Dolors
  • Quesada Diez, Rita
  • Berjano, Enrique J.
  • Grande Posa, Luís
  • Burdío Pinilla, Fernando
Radiofrequency energy has been used both experimentally and clinically to manage the pancreatic remnant after distal pancreatectomies. Our goal was to determine whether endoluminal radiofrequency (RF) ablation of the main pancreatic duct in large animals would be more efficient than glue occlusion as an exocrine pancreatic atrophy-inducing procedure. Thirty-four Landrace pigs were assigned to either the transpapilar (n = 16) or transection (n = 18) groups. The transection implied the pancreas neck was severed. In each of these groups the remaining distal pancreatic duct was occluded either by RF or by glue. In the transpapilar group complete atrophy was observed in all the RF cases, while atrophy was incomplete in all the members of the glue subgroup. The failure rate of the main pancreatic duct (usually expressed by a pseudocyst) in the transection groups was dramatically higher in the glue subgroup than the RF subgroups (9 out of 9 and 1 out of 9, respectively) and postoperative mortality occurred only in the glue subgroup (3 out of 9). These results show the superiority of endoluminal RF ablation over glue for main pancreatic duct occlusion, as seen by the degree of atrophy and fewer postoperative pancreatic fistulas., This work was supported by the Spanish Ministerio de Economía, Industria y Competitividad under the “Plan Estatal de Investigación, Desarrollo e Innovación Orientada a los Retos de la Sociedad” Grant TEC2014–52383–C3–R (TEC2014–52383–C3–1–R and TEC2014–52383–C3–3–R). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015–073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain).
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Development of a catheter-based technique for endoluminal radiofrequency sealing of pancreatic duct

  • Ewertowska, Elzbieta
  • Andaluz, Anna
  • Moll, Xavier
  • Adrià Aguilar
  • Garcia, Felix
  • Fondevila, Dolors
  • Quesada, Rita
  • Burdio, Fernando
  • Trujillo Guillen, Macarena|||0000-0003-4145-2188
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Introduction: Endoluminal sealing of the pancreatic duct by glue or sutures facilitates the management of the pancreatic stump. Our objective was to develop a catheter-based alternative for endoluminal radiofrequency (RF) sealing of the pancreatic duct. Materials and methods: We devised a novel RF ablation technique based on impedance-guided catheter pullback. First, bench tests were performed on ex vivo models to tune up the technique before the in vivo study, after which endoluminal RF sealing of a similar to 10 cm non-transected pancreatic duct was conducted on porcine models using a 3 Fr catheter. After 30 days, sealing effectiveness was assessed by a permeability test and a histological analysis. Results: The RF technique was feasible in all cases and delivered similar to 5 W of power on an initial impedance of 308 +/- 60 ohm. Electrical impedance evolution was similar in all cases and provided guidance for modulating the pullback speed to avoid tissue sticking and achieve a continuous lesion. During the follow-up the animals rate of weight gain was significantly reduced (p < 0.05). Apart from signs of exocrine atrophy, no other postoperative complications were found. At necropsy, the permeability test failed and the catheter could not be reintroduced endoluminally, confirming that sealing had been successful. The histological analysis revealed a homogeneous exocrine atrophy along the ablated segment in all the animals. Conclusions: Catheter-based RF ablation could be used effectively and safely for endoluminal sealing of the pancreatic duct. The findings suggest that a fully continuous lesion may not be required to obtain complete exocrine atrophy., This work was supported by the Spanish Ministerio de Economia, Industria y Competitividad under "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" (Grant TEC2014-52383-C3-R) and by the Spanish Ministerio de Ciencia, Innovacion y Universidades under "Programa Estatal de I+D+i Orientada a los Retos de la Sociedad" (Grant: RTI2018-094357-B-C21 and -C22). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015-073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain). Also funding was received from Secretaria de Estado de Investigacion, Desarrollo e Innovacion.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Numerical analysis of thermal impact of intramyocardial capillary blood flow during radiofrequency cardiac ablation

  • Pérez, Juan J|||0000-0001-8486-8699
  • González-Suárez, Ana|||0000-0002-1813-4176
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Purpose: The thermal effect of the intramyocardial blood perfusion on the size of lesions created by radiofrequency cardiac ablation (RFCA) has not been adequately studied to date. Our objective was to assess the impact of including this phenomenon in RFCA computer modelling in terms of the thermal lesion depth created.Methods: A computer model was built and computer simulations were conducted to assess the effect of including the blood perfusion term in the bioheat equation. This term mimics the intramyocardial blood flow (i.e., blood perfusion) in the cardiac wall at the site at which the RFCA is being conducted and hence represents a heat removing mechanism. When considered, blood perfusion rates ranged from 609 to 1719ml/min/kg. Two electrode design and modes were considered: a non-irrigated electrode with constant temperature mode and an irrigated electrode with constant power mode.Results: All the depths computed without including the blood perfusion term were larger than those that did include it, regardless of perfusion rate. The differences in lesion depth between ignoring and including blood perfusion increased over time; for a 60s RFCA they were 0.45 and 1mm for minimum and maximum perfusion rate, respectively. The differences were more or less independent of blood flow in the cardiac chamber, electrode type and ablation mode.Conclusions: The findings suggest that the heat-sink effect of blood perfusion should be taken into account in the case of ablations (>1minute) such as those conducted in RFCA of the ventricular wall., This work was supported by the Ministerio de Economia y Competitividad under "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" Grant "TEC2014- 52383-C3 [TEC2014-52383-C3-1-R]". A. Gonzalez-Suarez had a Postdoctoral Grant [APOSTD/2016/045] from the Generalitat Valenciana (Spain) and currently has a Postdoctoral Grant "Juan de la Cierva-formacion" [FJCI-2015-27202] supported by the Spanish Ministerio de Economia, Industria y Competitividad.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Effect of applied voltage, duration and repetition frequency of RF pulses for pain relief on temperature spikes and electrical field: A computer modeling study

  • Ewertowska, Elzbieta
  • Mercadal, Borja
  • Muñoz, Víctor
  • Ivorra, Antoni
  • Trujillo Guillen, Macarena|||0000-0003-4145-2188
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Background: The thermal and electrical effects of pulsed radiofrequency (PRF) for pain relief can be controlled by modifying the characteristics of the RF pulses applied. Our goal was to evaluate the influence of such modifications on the thermal and electric performance in tissue. Methods: A computational model was developed to compare the temperature and electric field time courses in tissue between a standard clinical protocol (45 V pulses, 20ms duration, 2 Hz repetition frequency) and a new protocol (55 V pulses, 5ms duration, 5 Hz repetition frequency) with a higher applied electric field but a smaller impact on temperature alterations in tissue. The effect of including a temperature controller was assessed. Complementarily, an agar-based experimental model was developed to validate the methodology employed in the computer modelling. Results: The new protocol increased the electric field magnitude reached in the tissue by around +20%, without increasing the temperature. The temperature controller was found to be the fundamental factor in avoiding thermal damage to the tissue and reduced the total number of pulses delivered by around 67%. The experimental results matched moderately well with those obtained from a computer model built especially to mimic the experimental conditions. Conclusions: For the same delivered energy, the new protocol significantly increases the magnitude of the applied electric field, which may be the reason why it is clinically more effective in achieving pain relief., This work was supported by the Spanish "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" under Grant TEC2014-52383-C3-R (TEC2014-52383-C3-1-R and TEC2014-52383-C3-2-R). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015-073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain).
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Computer modeling for radiofrequency bipolar ablation inside ducts and vessels: Relation between pullback speed and impedance progress

  • Pérez, Juan J|||0000-0001-8486-8699
  • Berjano, Enrique|||0000-0002-3247-2665
  • Ewertowska, Elzbieta
[EN] Background and Objectives Radiofrequency (RF)-induced ablation can be carried out inside ducts and vessels by simultaneously dragging a bipolar catheter while applying RF power. Our objective was to characterize the relation between pullback speed, impedance progress, and temperature distribution. Study Design/Materials and Methods We built a numerical model including a bipolar catheter, which is dragged inside a duct while RF power is applied between a pair of electrodes. The model solved a triple-coupled electrical, thermal, and mechanical problem. Lesions were assessed by an Arrhenius model. The numerical model's thermal and electrical characteristics were chosen to obtain the same initial impedance value as in the experiments: 560 omega at 16 degrees C (sample temperature). Results The catheter initially remained still, and the impedance was falling during the application of power. When pullback speed was too slow (<0.4 mm/s) impedance continued to drop when the catheter began to move, creating deep lesions, overheating and impedance roll-off, while at the faster speed (0.4-1.0 mm/s) impedance first rose slightly and then reached a plateau. There was a strong inverse relation between pullback speed and lesion depth. The hottest point was always around the second electrode, creating a kind of hot wake. Conclusions These findings confirm the close relationship between pullback speed and impedance progress, and suggest that the latter factor could be used to guide the procedure and achieve effective and safe ablations along the inner path of a duct or vessel., Spanish Ministerio de Ciencia, Innovacion y Universidades under "Programa Estatal de I + D + i Orientada a los Retos de la Sociedad; Contract grant number: RTI2018-094357-B-C21; Contract grant sponsor: Predoctoral Grant from Ministry of Economy, Industry and Competitiveness (Government of Spain); Contract grant number: BES-2015-073285; Contract grant sponsor: United States Air Force Office of Scientific Research.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Endoluminal radiofrequency ablation of the main pancreatic duct is a secure and effective method to produce pancreatic atrophy and to achieve stump closure

  • Andaluz, Anna
  • Ewertowska, Elzbieta
  • Moll, Xavier
  • Aguilar, Adrià
  • García, Félix
  • Fondevila, Dolors
  • Quesada, Rita
  • Grande, Luís
  • Burdío, Fernando
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Radiofrequency energy has been used both experimentally and clinically to manage the pancreatic remnant after distal pancreatectomies. Our goal was to determine whether endoluminal radiofrequency (RF) ablation of the main pancreatic duct in large animals would be more efficient than glue occlusion as an exocrine pancreatic atrophy-inducing procedure. Thirty-four Landrace pigs were assigned to either the transpapilar (n = 16) or transection (n = 18) groups. The transection implied the pancreas neck was severed. In each of these groups the remaining distal pancreatic duct was occluded either by RF or by glue. In the transpapilar group complete atrophy was observed in all the RF cases, while atrophy was incomplete in all the members of the glue subgroup. The failure rate of the main pancreatic duct (usually expressed by a pseudocyst) in the transection groups was dramatically higher in the glue subgroup than the RF subgroups (9 out of 9 and 1 out of 9, respectively) and postoperative mortality occurred only in the glue subgroup (3 out of 9). These results show the superiority of endoluminal RF ablation over glue for main pancreatic duct occlusion, as seen by the degree of atrophy and fewer postoperative pancreatic fistulas., This work was supported by the Spanish Ministerio de Economia, Industria y Competitividad under the "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" Grant TEC2014-52383-C3-R (TEC2014-52383-C3-1-R and TEC2014-52383-C3-3-R). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015-073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain).
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

A clinically-oriented computer model for radiofrequency ablation of hepatic tissue with internally cooled wet electrode

  • Ewertowska, Elzbieta
  • Quesada-Diez, Rita
  • Radosevic, A.
  • Andaluz, Anna
  • Moll, Xavier
  • García Arnás, F.
  • Burdío Pinilla, Fernando
  • Berjano, Enrique|||0000-0002-3247-2665
  • Trujillo Guillen, Macarena|||0000-0003-4145-2188
[EN] Purpose: To improve the computer modelling of radiofrequency ablation (RFA) by internally cooled wet (ICW) electrodes with added clinically oriented features. Methods: An improved RFA computer model by ICW electrode included: (1) a realistic spatial distribution of the infused saline, and (2) different domains to distinguish between healthy tissue, saline-infused tumour, and non-infused tumour, under the assumption that infused saline is retained within the tumour boundary. A realistic saline spatial distribution was obtained from an in vivo pig liver study. The computer results were analysed in terms of impedance evolution and coagulation zone (CZ) size, and were compared to the results of clinical trials conducted on 17 patients with the same ICW electrode. Results: The new features added to the model provided computer results that matched well with the clinical results. No roll-offs occurred during the 4-min ablation. CZ transversal diameter (4.10 +/- 0.19 cm) was similar to the computed diameter (4.16 cm). Including the tumour and saline infusion in the model involved (1) a reduction of the initial impedance by 10 - 20 ohm, (2) a delay in roll-off of 20 s and 70 - 100 s, respectively, and (3) 18 - 31% and 22 - 36% larger CZ size, respectively. The saline spatial distribution geometry was also seen to affect roll-off delay and CZ size. Conclusions: Using a three-compartment model and a realistic saline spatial distribution notably improves the match with the outcome of the clinical trials., This work was supported by the Spanish Ministerio de Economia, Industria y Competitividad under the "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" Grant TEC2014-52383-C3-R (TEC2014-52383-C3-1-R and TEC2014-52383-C3-3-R). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015-073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain).
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Endoluminal radiofrequency ablation of the main pancreatic duct is a secure and effective method to produce pancreatic atrophy and to achieve stump closure

  • Andaluz, Anna
  • Ewertowska, Elżbieta
  • Moll, Xavier
  • Aguilar, Adrià
  • García Arnas, Félix
  • Fondevila, Dolors
  • Quesada Diez, Rita
  • Berjano, Enrique J.
  • Grande Posa, Luís
  • Burdío Pinilla, Fernando
Radiofrequency energy has been used both experimentally and clinically to manage the pancreatic remnant after distal pancreatectomies. Our goal was to determine whether endoluminal radiofrequency (RF) ablation of the main pancreatic duct in large animals would be more efficient than glue occlusion as an exocrine pancreatic atrophy-inducing procedure. Thirty-four Landrace pigs were assigned to either the transpapilar (n = 16) or transection (n = 18) groups. The transection implied the pancreas neck was severed. In each of these groups the remaining distal pancreatic duct was occluded either by RF or by glue. In the transpapilar group complete atrophy was observed in all the RF cases, while atrophy was incomplete in all the members of the glue subgroup. The failure rate of the main pancreatic duct (usually expressed by a pseudocyst) in the transection groups was dramatically higher in the glue subgroup than the RF subgroups (9 out of 9 and 1 out of 9, respectively) and postoperative mortality occurred only in the glue subgroup (3 out of 9). These results show the superiority of endoluminal RF ablation over glue for main pancreatic duct occlusion, as seen by the degree of atrophy and fewer postoperative pancreatic fistulas., This work was supported by the Spanish Ministerio de Economía, Industria y Competitividad under the “Plan Estatal de Investigación, Desarrollo e Innovación Orientada a los Retos de la Sociedad” Grant TEC2014–52383–C3–R (TEC2014–52383–C3–1–R and TEC2014–52383–C3–3–R). Elzbieta Ewertowska has a Predoctoral Grant (BES-2015–073285) from the Ministry of Economy, Industry and Competitiveness (Government of Spain).
Dipòsit Digital de Documents de la UAB
Publicación Artículo científico (JournalArticle)

A clinically oriented computer model for radiofrequency ablation of hepatic tissue with internally cooled wet electrode

  • Ewertowska, Elzbieta
  • Quesada, R.|||0000-0002-8264-0356
  • Radosevic, Aleksandar
  • Andaluz Martínez, Anna|||0000-0001-8097-8110
  • Moll, Xavier|||0000-0002-2992-9361
  • García Arnas, Félix.|||0000-0001-6651-3173
  • Berjano, Enrique|||0000-0002-3247-2665
  • Burdío, Fernando|||0000-0003-3038-0086
  • Trujillo Guillén, Macarena|||0000-0003-4145-2188
Purpose: To improve the computer modelling of radiofrequency ablation (RFA) by internally cooled wet (ICW) electrodes with added clinically oriented features. Methods: An improved RFA computer model by ICW electrode included: (1) a realistic spatial distribution of the infused saline, and (2) different domains to distinguish between healthy tissue, saline-infused tumour, and non-infused tumour, under the assumption that infused saline is retained within the tumour boundary. A realistic saline spatial distribution was obtained from an in vivo pig liver study. The computer results were analysed in terms of impedance evolution and coagulation zone (CZ) size, and were compared to the results of clinical trials conducted on 17 patients with the same ICW electrode. Results: The new features added to the model provided computer results that matched well with the clinical results. No roll-offs occurred during the 4-min ablation. CZ transversal diameter (4.10 ± 0.19 cm) was similar to the computed diameter (4.16 cm). Including the tumour and saline infusion in the model involved (1) a reduction of the initial impedance by 10 - 20 Ω, (2) a delay in roll-off of 20 s and 70 - 100 s, respectively, and (3) 18 - 31% and 22 - 36% larger CZ size, respectively. The saline spatial distribution geometry was also seen to affect roll-off delay and CZ size. Conclusions: Using a three-compartment model and a realistic saline spatial distribution notably improves the match with the outcome of the clinical trials.
Dipòsit Digital de Documents de la UAB
Publicación Artículo científico (JournalArticle)

Development of a catheter-based technique for endoluminal radiofrequency sealing of pancreatic duct

  • Ewertowska, Elzbieta
  • Andaluz Martínez, Anna|||0000-0001-8097-8110
  • Moll, Xavier|||0000-0002-2992-9361
  • Aguilar Catalán, Adrià|||0000-0003-2403-9383
  • García Arnas, Félix|||0000-0001-6651-3173
  • Fondevila, Dolors|||0000-0001-7443-0175
  • Quesada, R.|||0000-0002-8264-0356
  • Trujillo Guillén, Macarena|||0000-0003-4145-2188
  • Burdío, Fernando|||0000-0003-3038-0086
  • Berjano, Enrique|||0000-0002-3247-2665
Introduction: Endoluminal sealing of the pancreatic duct by glue or sutures facilitates the management of the pancreatic stump. Our objective was to develop a catheter-based alternative for endoluminal radiofrequency (RF) sealing of the pancreatic duct. Materials and methods: We devised a novel RF ablation technique based on impedance-guided catheter pullback. First, bench tests were performed on ex vivo models to tune up the technique before the in vivo study, after which endoluminal RF sealing of a ∼10 cm non-transected pancreatic duct was conducted on porcine models using a 3 Fr catheter. After 30 days, sealing effectiveness was assessed by a permeability test and a histological analysis. Results: The RF technique was feasible in all cases and delivered ∼5 W of power on an initial impedance of 308 ± 60 Ω. Electrical impedance evolution was similar in all cases and provided guidance for modulating the pullback speed to avoid tissue sticking and achieve a continuous lesion. During the follow-up the animals rate of weight gain was significantly reduced (p < 0.05). Apart from signs of exocrine atrophy, no other postoperative complications were found. At necropsy, the permeability test failed and the catheter could not be reintroduced endoluminally, confirming that sealing had been successful. The histological analysis revealed a homogeneous exocrine atrophy along the ablated segment in all the animals. Conclusions: Catheter-based RF ablation could be used effectively and safely for endoluminal sealing of the pancreatic duct. The findings suggest that a fully continuous lesion may not be required to obtain complete exocrine atrophy.