Evaluation of the two-voltage method for parallel-plate ionization chambers irradiated with pulsed beams
Air-vented ionization chambers exposed to clinical radiation beams may suffer from recombination during the drift of the charge carriers towards the electrodes. Thus, dosimetry protocols recommend the use of a correction factor, usually denominated saturation factor ( ksat$k_{\rm sat}$ ), to correct the ionization chamber readout for the incomplete collection of charge. The two-voltage method (TVM) is the recommended methodology for the calculation of the saturation factor, however, it is based on the early Boag model, which only takes into account the presence of positive and negative ions in the ionization chamber and does not account for the electric field screening or the free electron contribution to the signal., Jose Paz-Martín has received a predoctoral research contract from the Xunta de Galicia regional govern-ment. Juan Pardo-Montero acknowledges the support of Xunta de Galicia, Axencia Galega de Innovación (grant IN607D2022). This work has been developed under the project 18HLT04 UHDpulse with funding from the EMPIR innovation program. This work has received funding from ‘la Caixa’ Foundation under the grant agreement HR23-00718 (‘Dosimetry monitor for FLASH therapy’ project) and from grant PLEC2022-009476 funded by MCIN/AEI/10.13039/501100011033 and by the ‘European Union NextGenerationEU/PRTR’. The authors thank Christoph Makowski for the electron linear accelerator maintenance during the measurement campaign, Olaf Tappe for the machining work, and Thomas Hackel for the alanine measurements., Peer reviewed