Publication Tesis de maestría (MasterThesis)

Characterization and simulation of a CdTe detector for use in PET

  • Ariño-Estrada, Gerard|||0000-0002-6411-191X
The Voxel Imaging PET (VIP) Path nder project got the 4 year European Research Council FP7 grant in 2010 to prove the feasibility of using CdTe detectors in a novel conceptual design of PET scanner. The work presented in this thesis is a part of the VIP project and consists of, on the one hand, the characterization of a CdTe detector in terms of energy resolution and coincidence time resolution and, on the other hand, the simulation of the setup with the single detector in order to extend the results to the full PET scanner. An energy resolution of 0.98% at 511 keV with a bias voltage of 1000 V/mm has been measured at low temperature T=-8 ºC. The coincidence time distribution of two twin detectors has been found to be as low as 6 ns FWHM for events with energies above 500 keV under the same temperature and bias conditions. The measured energy and time resolution values are compatible with similar ndings available in the literature and prove the excellent potential of CdTe for PET applications. This results have been presented in form of a poster contribution at the IEEE NSS/MIC & RTSD 2011 conference in October 2011 in Valencia and at the iWoRID 2012 conference in July 2012 in Coimbra, Portugal. They have been also submitted for publication to "Journal of Instrumentation (JINST)" in September 2012.
Dipòsit Digital de Documents de la UAB
Publication Artículo científico (JournalArticle)

Simulation of charge transport in pixelated CdTe

  • Kolstein, Machiel|||0000-0002-5482-6743
  • Ariño-Estrada, Gerard|||0000-0002-6411-191X
  • Chmeissani, Mokhtar|||0000-0002-2287-4791
  • De Lorenzo, Gianluca|||0000-0003-1795-0988
The Voxel Imaging PET (VIP) Pathfinder project intends to show the advantages of using pixelated semiconductor technology for nuclear medicine applications to achieve an improved image reconstruction without efficiency loss. It proposes designs for Positron Emission Tomography (PET), Positron Emission Mammography (PEM) and Compton gamma camera detectors with a large number of signal channels (of the order of 10 6). The design is based on the use of a pixelated CdTe Schottky detector to have optimal energy and spatial resolution. An individual read-out channel is dedicated for each detector voxel of size 1 × 1 × 2 mm 3 using an application-specific integrated circuit (ASIC) which the VIP project has designed, developed and is currently evaluating experimentally. The behaviour of the signal charge carriers in CdTe should be well understood because it has an impact on the performance of the readout channels. For this purpose the Finite Element Method (FEM) Multiphysics COMSOL software package has been used to simulate the behaviour of signal charge carriers in CdTe and extract values for the expected charge sharing depending on the impact point and bias voltage. The results on charge sharing obtained with COMSOL are combined with GAMOS, a Geant based particle tracking Monte Carlo software package, to get a full evaluation of the amount of charge sharing in pixelated CdTe for different gamma impact points.
Dipòsit Digital de Documents de la UAB
Publication Artículo científico (JournalArticle)

Measurement of mobility and lifetime of electrons and holes in a Schottky CdTe diode

  • Ariño-Estrada, Gerard|||0000-0002-6411-191X
  • Chmeissani, Mokhtar|||0000-0002-2287-4791
  • De Lorenzo, Gianluca|||0000-0003-1795-0988
  • Kolstein, Machiel|||0000-0002-5482-6743
  • Puigdengoles, Carles|||0000-0002-1372-5452
  • García, J.
  • Cabruja i Casas, Enric|||0000-0003-4505-6465
We report on the measurement of drift properties of electrons and holes in a CdTe diode grown by the travelling heating method (THM). Mobility and lifetime of both charge carriers has been measured independently at room temperature and fixed bias voltage using charge integration readout electronics. Both electrode sides of the detector have been exposed to a 241 Am source in order to obtain events with full contributions of either electrons or holes. The drift time has been measured to obtain the mobility for each charge carrier. The Hecht equation has been employed to evaluate the lifetime. The measured values for μτ (mobility-lifetime product) are in agreement with earlier published data.