The AX-PET demonstrator - Design, construction and characterization

  • Beltrame, P.
  • Bolle, E.
  • Braem, A.
  • Casella, C.
  • Chesi, E.
  • Clinthorne, N.
  • De Leo, R.
  • Dissertori, G.
  • Djambazov, L.
  • Fanti, V.
  • Heller, M.
  • Joram, C.
  • Kagan, H.
  • Lustermann, W.
  • Meddi, F.
  • Nappi, E.
  • Nessi-Tedaldi, F.
  • Oliver, Josep F.
  • Pauss, F.
  • Rafecas, Magdalena
  • Renker, D.
  • Rudge, A.
  • Schinzel, D.
  • Schneider, T.
  • Séguinot, J.
  • Solevi, P.
  • Stapnes, S.
  • Weilhammer, P.
Axial PET is a novel geometrical concept for Positron Emission Tomography (PET), based on layers of long scintillating crystals axially aligned with the bore axis. The axial coordinate is obtained from arrays of wavelength shifting (WLS) plastic strips placed orthogonally to the crystals. This article describes the design, construction and performance evaluation of a demonstrator set-up which consists of two identical detector modules, used in coincidence. Each module comprises 48 LYSO crystals of 100 mm length and 156 WLS strips. Crystals and strips are readout by Geiger-mode Avalanche Photo Diodes (G-APDs). The signals from the two modules are processed by fully analog front-end electronics and recorded in coincidence by a VME-based data acquisition system. Measurements with point-like <sup>22</sup>Na sources, with the modules used both individually and in coincidence mode, allowed for a complete performance evaluation up to the focal plane reconstruction of point sources. The results obtained are in good agreement with expectations and proved the set-up to be ready for the next evaluation phase with PET phantoms filled with radiotracers. © 2011 Elsevier B.V. All right reserved., The work is partially funded through the Spanish Ministry of Science & Innovation under Grant FPA2008-02419-E and project Grant TEC2007-61047. P. Solevi is supported by the Marie Curie Intra-European Fellowship Grant 237620 for the project INSPET. P. Beltrame is supported through the Marie Curie Initial Training Network MC-PAD (GA no. 214560-2)., No