RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Coincidence summing correction factors for 238U and 232Th decay series using the Monte Carlo method

  • Ordóñez-Ródenas, José|||0000-0001-8150-4873
  • Gallardo Bermell, Sergio|||0000-0002-3703-9983
  • Martorell Alsina, Sebastián Salvador|||0000-0003-1706-4740
  • Ortiz Moragón, Josefina
[EN] Environmental samples analyzed in gamma spectrometry laboratories usually contain natural radionuclides such as 238U and 232Th. Using gamma spectrometry techniques is possible to estimate the activity of these radionuclides by measuring the gamma emissions of radionuclides belonging to their decay chain. Nonetheless, some of these radionuclides emit photons in cascade presenting Coincidence Summing (CS), which if not corrected, may affect the final activity quantification. The aim of this work is to apply the Monte Carlo method to calculate the True Summing Correction Factors (TSCFs) for 238U and 232Th decay series for different sample configurations (geometry and matrix) using the GEANT4 toolkit. In order to validate the results provided by GEANT4 using the RDM, the software TRUECOINC has been applied to calculate also the TSCFs. In addition, the influence of the geometry/matrix on the TSCFs is analyzed., The authors gratefully acknowledge financial support from the Catedra CSN-UPV Vicente Serradell, Spain as well as the Laboratorio de Radiactividad Ambiental (Universitat Politecnica de Valencia), Spain for the dedicated funding and resources to this research work under Grant no. FPI-2015-S2-1576
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Intercomparison of full energy peak efficiency curves for an HPGe detector using MCNP6 and GEANT4

  • Ordóñez-Ródenas, José|||0000-0001-8150-4873
  • Gallardo Bermell, Sergio|||0000-0002-3703-9983
  • Sáez-Muñoz, Marina|||0000-0003-0180-7167
  • Martorell Alsina, Sebastián Salvador|||0000-0003-1706-4740
  • Ortiz Moragón, Josefina
[EN] This work is focused on studying the capabilities of different Monte Carlo tools to complement the Full Energy Peak Efficiency (FEPE) calibration procedures of the Environmental Radioactivity Laboratory of the Universitat Politècnica de València, UPV. In this frame, detailed models of a High Purity Germanium detector have been implemented using MCNP6 and GEANT4. Accordingly, an inter-comparison with experimental values ensures the model validation and improves the analysis of the physics phenomena involved. The validation of the models is determined by a quantitative comparison between the simulated and the measured efficiencies over the energy range under study (59.54¿1836.01 keV). The results show discrepancies between both Monte Carlo tools for 139Ce, 88Y and 60Co as GEANT4 is able to simulate the coincidence summing effect of these radionuclides., The authors gratefully acknowledge the Catedra CSN-UPV Vicente Serradell, Spain as well as the Laboratorio de Radiactividad Ambiental (Universitat Politecnica de Valencia), Spain for the dedicated funding and resources to this research work under Grant no. FPI-2015-S2-1576.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Monte Carlo model of a BEGe detector to support gamma-spectrometry in an emergency response

  • Ordóñez-Ródenas, José|||0000-0001-8150-4873
  • Martorell Alsina, Sebastián Salvador|||0000-0003-1706-4740
  • Gallardo Bermell, Sergio|||0000-0002-3703-9983
  • Ortiz Moragón, Josefina
[EN] One of the problems of measuring high radioactivity in an emergency scenario is the fact that the detector can become saturated or reach a measuring dead time too high to give reliable results, which means repeating the measurement in different conditions with the associated delay in obtaining the results and the laboratory workers' risk of exposure. The counting rate can be controlled by varying the sample-to-detector distance as well as by using different source volumes. A Monte Carlo model of a BEGe detector was developed to analyse the system efficiency response for several measuring configurations (distances and volumes) using the MCNP6 code. The total efficiency curves were obtained for an energy range between 59.5 keV and 1836 keV. The simulations provided an estimation of the admissible gamma-rate for different source volumes (in water matrix) and sample-to-detector distances to avoid detector saturation in given measurement conditions. The results were a compromise between geometry, distance and measuring time for certain emergency situations. Three case studies are provided to show the approach's effectiveness., The authors gratefully acknowledge the Catedra CSN-UPV Vicente Serradell, Spain as well as the Laboratorio de Radiactividad Ambiental (Universitat Politecnica de Valencia), Spain, for the dedicated funding and resources for this research work under Grant No. FPI-2015-S2-1576. The authors are also grateful to the Valencian Agency for Security and Emergency and the Generalitat Valenciana for their support in the "Convenio de colaboracion entre la Agencia Valenciana de Seguridad y Respuesta a las Emergencias y la Universitat Politecnica de Valencia para el Desarrollo del Plan de Vigilancia Radiologica en Emergencias" under Grant No. S7042000, 2018.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Contribución de congreso (ConferenceOutput)

Aplicación de un modelo de detector BEGe en MCNP6 como apoyo a la realización de medidas por espectrometría gamma en un laboratorio en caso de emergencia nuclear o radiactiva

  • Ordóñez-Ródenas, José|||0000-0001-8150-4873
  • Martorell Alsina, Sebastián Salvador|||0000-0003-1706-4740
  • Gallardo Bermell, Sergio|||0000-0002-3703-9983
  • Ortiz Moragón, Josefina
[ES] Uno de los problemas al medir actividades altas como las de un escenario de emergencia es la posible saturación del detector o la obtención un tiempo muerto de medición demasiado alto como para dar resultados fiables. Esto llevaría a repetir la medida en diferentes condiciones aumentando el tiempo de respuesta y, por consiguiente, el riesgo de exposición de los trabajadores del laboratorio. La tasa de cuentas medidas por el detector se puede controlar variando la distancia de la muestra al detector, así como utilizando diferentes volúmenes de fuente. Se ha desarrollado un modelo de Monte Carlo de un detector BEGe para analizar la respuesta en eficiencia del sistema para varias geometrías de medición (distancias y volúmenes). Las curvas de eficiencia total para muchas de las configuraciones utilizadas en el laboratorio se obtienen utilizando el código MCNP6 para un rango de energía de entre 59.5 keV y 1836.0 keV. Las simulaciones han permitido estimar la emisión gamma admisible para diferentes volúmenes de fuente (en matriz de agua) y distancias de muestra a detector que evite la saturación del mismo. Los resultados conducen a lograr un compromiso entre la geometría, la distancia y el tiempo de medición para una determinada condición en situaciones de emergencia. Se proporcionan tres casos de aplicación para explicar cómo funciona el enfoque propuesto., Los autores agradecen a la Cátedra CSN-UPV Vicente Serradell, así como al Laboratorio de Radiactividad Ambiental (Universitat Politècnica de València), por la financiación y los recursos dedicados a este trabajo de investigación en virtud de la subvención no. FPI-2015-S2-1576. Además, los autores agradecen a la Agencia Valenciana de Seguridad y Emergencias y a la Generalitat Valenciana por su apoyo en el "Convenio de colaboración entre la Agencia Valenciana de Seguridad y Respuesta a las Emergencias y la Universitat Politècnica de València para el Desarrollo del Plan de Vigilancia Radiológica en Emergencias¿ bajo la Subvención No. S7042000, 2018.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Contribución de congreso (ConferenceOutput)

Estudio de la respuesta de un TLD100-HARSHAW mediante el código de Monte Carlo MCNP6

  • Ordóñez-Ródenas, José|||0000-0001-8150-4873
  • Gallardo Bermell, Sergio|||0000-0002-3703-9983
  • Martorell Alsina, Sebastián Salvador|||0000-0003-1706-4740
  • Ortiz Moragón, Josefina
[ES] El código MCNP6 se ha utilizado para desarrollar un modelo de Monte Carlo del TLD100- HARSHAW, incluyendo una fuente colimada y un fantoma de agua con paredes de PMMA (ISO water slab phantom). Para calcular el kerma en aire y las dosis superficial Hp (0.07) y profunda Hp (10), se aplica un tally F6 (MeV/g). Este modelo permite obtener la función de respuesta relativa en energía (RRE) normalizada al 137Cs para haces de diferente calidad y ángulos de incidencia. La validación del modelo se lleva a cabo comparando datos simulados con experimentales utilizando diferentes calidades de haz de rayos-X de acuerdo con la norma ISO-4037-3. El código MCNP6 representa una herramienta muy útil para analizar las mediciones experimentales y a su vez para realizar calibraciones de TLDs sin necesidad de irradiarlos., Los autores agradecen a la Cátedra CSN-UPV Vicente Serradell, así como al Laboratorio de Radiactividad Ambiental (Universitat Politècnica de València), por la financiación y los recursos dedicados a este trabajo de investigación en virtud de la subvención no. FPI-2015-S2-1576.