Measuring specific retention volumes in capillary gas chromatography with improved accuracy and precision

  • Quintanilla-López, Jesús Eduardo
  • Lebrón-Aguilar, Rosa
  • Tello, Ana María
  • García-Domínguez, José Antonio
9 pages, 2 figures, 7 tables, The specific retention volume (V(g)), which has traditionally been determined in gas chromatography using long, heavily loaded packed columns, should produce better, more reliable values if capillary columns are used. However, the smaller dimensions of some of the parameters involved introduces additional sources of error which have been overcome with a recently proposed equation. A careful consideration of the sources of error involved has been carried out using both the traditional and the new equations. The conclusion drawn is that the new equation will produce reliable V(g) values with expected errors (1%) lower than those expected from packed columns (2%) using the traditional equation., This work was carried out with the financial assistance of the Dirección General de Investigación Cientifica y Técnica of Spain, under Projects PB91-0077 and PB94-0218. We are particularly grateful to Q. Dai, who provided data on the density of the stationary phase with carefully performed experiments., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

The hold-up time in gas chromatography. III. Linearity of the plot of in t(R)/' of the n-alkanes vs. carbon number questioned

  • García-Domínguez, José Antonio
  • Quintanilla-López, Jesús Eduardo
  • Lebrón-Aguilar, Rosa
6 pages, 5 figures, 1 table, Three arguments are presented to show that the plot of the logarithms of the adjusted retention times of n-alkanes eluted under isothermal conditions versus their chain length is not a straight line, as generally accepted, even for n-alkanes of seventeen carbon atoms. Comparison of values of t(R)/' deduced by methods based on the linearity assumption (In t(R)/'=mz+n) and on a different approach [t(R)=A+exp(B+Cz(D))] shows that experimental retention times do not fit the first expression, independently of the value used for the hold-up time, even if the correlation coefficient of the fit is better than 0.99999., The work presented in this paper has been carriederal out under Project PB94-0218 of the Direccion General de Investigacion Cientıfica y Tecnica (DGICYT) of Spain. The authors thank A.M. Tello, M.A. de Diago and A. Fernandez-Torres for their ´ help with some of the chromatograms and numerical calculations., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Hold-up time in gas chromatography. IV. Improved determination of Kovats' retention indices

  • Lebrón-Aguilar, Rosa
  • García-Domínguez, José Antonio
  • Quintanilla-López, Jesús Eduardo
8 pages, 2 pages, 4 tables, Kovats' retention indices of n-alkanes eluted under isothermal conditions, in nine columns (packed and capillary), at temperatures between 60 and 150°C are calculated by three different methods, and the results are compared with their theoretical values. Two of the methods are based on the linearity assumption of the plot of the logarithms of the adjusted retention times versus carbon number. The third one does not accept the linearity assumption. Results show that methods based on the linearity assumption may produce a clear deviation from the true value of the retention index, while the third method tested, considerably improves the accuracy of the results, showing a similar precision than that obtained by the classical procedures., The work presented in this paper has been carried out under Project PB94-0218 of the Direccion General de Investigacion Cientıfica y Tecnica ´ ´ ´ (DGICYT) of Spain., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Hold-up time in gas chromatography. V. Dependence of the retention of n-alkanes on the chromatographic variables in isothermal gas chromatography

  • Quintanilla-López, Jesús Eduardo
  • Lebrón-Aguilar, Rosa
  • García-Domínguez, José Antonio
11 pages, 4 figures, 2 tables, The chromatographic behaviour of n-alkanes and other homologous series in isothermal gas chromatography has been shown to depart from the "linear" representation of the logarithms of the adjusted retention times vs. carbon number. One of the expressions proposed to describe this behaviour is tR(z)=A+exp(B+CzD). In this paper, a regression analysis shows that three of the parameters of the equation depend on different chromatographic variables such as hold-up time, average linear gas velocity, volume and polarity of the stationary phase and temperature of the column. The fourth parameter (D), responsible for the departure from the "linearity", does not depend on any chromatographic variable, and represents a gradual decrease of the contribution of the methylene groups to the general properties of n-alkanes, with no relation to the chromatographic phenomenon., This work has been carried out with financial help from the Direccion General de Investigacion Cientıfica y Tecnica (DGICYT) of Spain, under projects PB94-0218 and PB97-1214. R.L.-A. is grateful to the Ministry of Work for a short duration contract, and J.E.Q.-L. thanks Consejo Superior de Investigaciones ´ Cientıficas (CSIC) for a grant., Peer reviewed