DIGITAL.CSIC. Repositorio Institucional del CSIC
Publicación
Artículo científico (JournalArticle)
Characterization of a Compton camera based on the TOFPET2 ASIC
The use of Compton cameras for medical imaging and its interest as a hadron therapy treatment monitoring has increased in the last decade with the development of silicon photomultipliers. MACACO is a Compton camera prototype designed and assembled at the IRIS group of IFIC-Valencia. This Compton camera is based on monolithic Lanthanum (III) Bromide crystals and silicon photomultipliers, and employs the novel TOFPET2 ASIC as readout electronics. This system emerged as an alternative to MACACO II prototype, with the aim of improving its limited time resolution. To test the performance of the ASIC in a Compton camera setup, the prototype was characterized, both in laboratory and in-beam. A time resolution of 1.5 ns was obtained after time corrections, which improves greatly the performance of the MACACO II. Moreover, the results obtained at high photon energies demonstrate the ability of the system to obtain 1 mm displacements of the reconstructed spots. The results reinforce the potential of the system as a monitoring device for hadron therapy., This work has received funding from the Generalitat Valenciana, Spain (AICO/2019070) and from the Spanish Ministerio de Ciencia e Innovación (PID2019-110657RB-100). Group members are supported by
UVEG Atracció de Talent, FPU, CDEIGENT, SEJIGENT and Generalitat Valenciana contracts. M.C. Jiménez-Ramos acknowledges the support to this work through a VI PPIT-US contract., Peer reviewed