Publicación Artículo científico (JournalArticle)

A computational comparison of radiofrequency and pulsed field ablation in terms of lesion morphology in the cardiac chamber

  • Gómez‑Barea, Mario
  • Garcia-Sanchez, Tomas
  • Ivorra Cano, Antoni, 1974-
Pulsed Field Ablation (PFA) has been developed over the last years as a novel electrical ablation technique for treating cardiac arrhythmias. It is based on irreversible electroporation which is a non-thermal phenomenon innocuous to the extracellular matrix and, because of that, PFA is considered to be safer than the reference technique, Radiofrequency Ablation (RFA). However, possible differences in lesion morphology between both techniques have been poorly studied. Simulations including electric, thermal and fluid physics were performed in a simplified model of the cardiac chamber which, in essence, consisted of a slab of myocardium with blood in motion on the top. Monopolar and bipolar catheter configurations were studied. Different blood velocities and catheter orientations were assayed. RFA was simulated assuming a conventional temperature-controlled approach. The PFA treatment was assumed to consist in a sequence of 20 biphasic bursts (100 µs duration). Simulations indicate that, for equivalent lesion depths, PFA lesions are wider, larger and more symmetrical than RFA lesions for both catheter configurations. RFA lesions display a great dependence on blood velocity while PFA lesions dependence is negligible on it. For the monopolar configuration, catheter angle with respect to the cardiac surface impacted both ablation techniques but in opposite sense. The orientation of the catheter with respect to blood flow direction only affected RFA lesions. In this study, substantial morphological differences between RFA and PFA lesions were predicted numerically. Negligible dependence of PFA on blood flow velocity and direction is a potential important advantage of this technique over RFA., This work was partially supported by project (PID2019-110120RBI00/AEI/10.13039/501100011033) from the Ministry of Science, Innovation and Universities and the State Research Agency of the Spanish government. Mario Gómez-Barea research is supported by a scholarship from the UPF. Tomás García-Sánchez acknowledges the funding received from the European Union’s Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie grant agreement No 892393 (TACAIRE). Antoni Ivorra gratefully acknowledges the financial support by ICREA under the ICREA Academia programme.
Repositorio Digital de la UPF
Publicación Artículo científico (JournalArticle)

Is a single lethal electric field threshold sufficient to characterize the lesion size in computational modeling of cardiac pulsed-field ablation?

  • Petras, Argyrios
  • Amorós Figueras, Gerard
  • Moreno Weidmann, Zoraida
  • García Sánchez, Tomás
  • Viladés-Medel, David
  • Ivorra Cano, Antoni, 1974-
  • Guerra, Jose M.
  • Gerardo-Giorda, Luca
Background: Pulsed-field ablation (PFA) is a novel cardiac ablation technology based on irreversible electroporation (IRE). PFA computational models rely on identification of a lethal electric field threshold to predict the IRE area. However, the predicted lesion anisotropy ratios (width over depth) vary extensively among recent studies, and these discrepancies remain a subject of discussion. Objective: This work aims to evaluate the predicted lesion anisotropy ratios using a PFA computational model by applying it to an open-chest in vivo porcine model geometry. Methods: Six domestic swine underwent epicardial PFA applications using a previously described waveform protocol. Animals were killed at least 3 hours after the last ablation, and lesions were assessed using triphenyltetrazolium chloride (TTC) staining. Numeric simulations were performed on a segmented and meshed porcine thoracic computed tomography (CT) scan, mimicking the open-chest experimental setup. Results: The maximum width of all simulated lesions was observed at the epicardial surface. The anisotropy ratios (AR) of the experimental lesions were smaller than the simulated ones (AR experimental vs simulated, 1.0-1.7 vs 2-2.7; Q1-Q3 quartiles). Increasing the peak voltage resulted in larger lesions; however, the computational model clearly underestimated the increase in lesion depth compared with the experimental data. Conclusion: Our computational model shows that a single lethal electric field threshold is insufficient to accurately predict both lesion depth and width in cardiac PFA. Our study suggests that for the given PFA waveforms, a threshold between 270 and 500 V/cm provides satisfactory lesion depth estimations, and a higher threshold between 790 and 1000 V/cm better captures the lesion width., A.P and L.GG acknowledge the partial support of the State of Upper Austria. The research was funded in part by the Austrian Science Fund (FWF) P35374N. For the purpose of Open Access, the author has applied a CC-BY public copyright license to any Author Accepted Manuscript (AAM) version arising from this submission. A.P acknowledges the FFG Bridge grant PFA-Mod. This work was partially supported by project (PID2019-110120RBI00 /AEI/10.13039/ 501100011033) from the Ministry of Science, Innovation and Universities and the State Research Agency of the Spanish government. This work was also partially supported by project ISCIII DTS20-00111 (Instituto de Salud Carlos III and cofunded by FEDER). This project has received funding from the European Union's Horizon 2020 research and innovation program under the Marie Sklodowska-Curie grant agreement No 892393 (TACAIRE).
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publicación Artículo científico (JournalArticle)

Dynamics of High-Density Unipolar Epicardial Electrograms During PFA

  • Amorós-Figueras, Gerard
  • Casabella, Sergi
  • Moreno-Weidmann, Zoraida
  • Ivorra Cano, Antoni
  • Guerra, Jose M.
  • García-Sánchez, Tomás
Pulsed field ablation (PFA) is a novel nonthermal cardiac ablation technology based on irreversible electroporation (IRE). While areas of IRE lead to durable lesions, the surrounding regions, where reversible electroporation occurs, recover. The behavior of local electrograms in areas of different electroporation levels remains unknown. The goal of this study is to characterize electrogram dynamics after PFA in IRE and reversible electroporation areas. METHODS: A total of 6 domestic swine were used. PFA was applied in the epicardium of the right and left ventricles using a focal monopolar catheter. Additional radiofrequency ablations were performed. Epicardial unipolar electrograms were acquired at baseline and for 60 minutes post PFA/radiofrequency ablation using a high-density electrode matrix attached to the epicardium. Electrogram dynamics were analyzed in areas corresponding to different levels of electroporation. Acute lesion formation was assessed after 3 to 5 hours by triphenyl tetrazolium chloride staining. RESULTS: Electrogram analysis demonstrated a clear association between electrogram changes and the level of electroporation. Immediately after PFA, electrograms displayed the following: a significant decrease in R/S-wave amplitude; a large elevation of the ST-segment; and a large decrease in their |(dV/dt)|. Marked changes in electrograms were observed beyond the lesion area. Thereafter, a gradual recovery was observed. The evolution of all the electrogram parameters throughout the 60 minutes after PFA was significantly different (P<0.05) between the IRE and reversible electroporation areas. Acute lesion staining showed significantly larger depth for PFA lesions compared with radiofrequency ablation. CONCLUSIONS: This study shows that unipolar electrograms can differentiate between reversible electroporation and IRE areas during the first 30 minutes post ablation. Differences after the first 30 minutes are less evident. Our findings could result useful for immediate lesion assessment after PFA and warrant further investigation., This work was partially supported by project (PID2019-110120RBI00/AEI/10.13039/501100011033) from the Ministry of Science, Innovation and Universities and the State Research Agency of the Spanish government. This work was also partially supported by project ISCIII DTS20-00111 (Instituto de Salud Carlos III and cofunded by FEDER). This project has received funding from the European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie grant agreement No 892393 (TACAIRE). Dr Ivorra gratefully acknowledges financial support by ICREA under the ICREA Academia program.
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

Is a single lethal electric field threshold sufficient to characterize the lesion size in computational modeling of cardiac pulsed-field ablation?

  • Petras, Argyrios
  • Amorós Figueras, Gerard
  • Moreno Weidmann, Zoraida
  • García Sánchez, Tomás
  • Viladés-Medel, David
  • Ivorra Cano, Antoni, 1974-
  • Guerra, Jose M.
  • Gerardo-Giorda, Luca
Background: Pulsed-field ablation (PFA) is a novel cardiac ablation technology based on irreversible electroporation (IRE). PFA computational models rely on identification of a lethal electric field threshold to predict the IRE area. However, the predicted lesion anisotropy ratios (width over depth) vary extensively among recent studies, and these discrepancies remain a subject of discussion. Objective: This work aims to evaluate the predicted lesion anisotropy ratios using a PFA computational model by applying it to an open-chest in vivo porcine model geometry. Methods: Six domestic swine underwent epicardial PFA applications using a previously described waveform protocol. Animals were killed at least 3 hours after the last ablation, and lesions were assessed using triphenyltetrazolium chloride (TTC) staining. Numeric simulations were performed on a segmented and meshed porcine thoracic computed tomography (CT) scan, mimicking the open-chest experimental setup. Results: The maximum width of all simulated lesions was observed at the epicardial surface. The anisotropy ratios (AR) of the experimental lesions were smaller than the simulated ones (AR experimental vs simulated, 1.0-1.7 vs 2-2.7; Q1-Q3 quartiles). Increasing the peak voltage resulted in larger lesions; however, the computational model clearly underestimated the increase in lesion depth compared with the experimental data. Conclusion: Our computational model shows that a single lethal electric field threshold is insufficient to accurately predict both lesion depth and width in cardiac PFA. Our study suggests that for the given PFA waveforms, a threshold between 270 and 500 V/cm provides satisfactory lesion depth estimations, and a higher threshold between 790 and 1000 V/cm better captures the lesion width., A.P and L.GG acknowledge the partial support of the State of Upper Austria. The research was funded in part by the Austrian Science Fund (FWF) P35374N. For the purpose of Open Access, the author has applied a CC-BY public copyright license to any Author Accepted Manuscript (AAM) version arising from this submission. A.P acknowledges the FFG Bridge grant PFA-Mod. This work was partially supported by project (PID2019-110120RBI00 /AEI/10.13039/ 501100011033) from the Ministry of Science, Innovation and Universities and the State Research Agency of the Spanish government. This work was also partially supported by project ISCIII DTS20-00111 (Instituto de Salud Carlos III and cofunded by FEDER). This project has received funding from the European Union's Horizon 2020 research and innovation program under the Marie Sklodowska-Curie grant agreement No 892393 (TACAIRE).
Recercat. Dipósit de la Recerca de Catalunya
Publicación Artículo científico (JournalArticle)

A computational comparison of radiofrequency and pulsed field ablation in terms of lesion morphology in the cardiac chamber

  • Gómez‑Barea, Mario
  • Garcia-Sanchez, Tomas
  • Ivorra Cano, Antoni, 1974-
Pulsed Field Ablation (PFA) has been developed over the last years as a novel electrical ablation technique for treating cardiac arrhythmias. It is based on irreversible electroporation which is a non-thermal phenomenon innocuous to the extracellular matrix and, because of that, PFA is considered to be safer than the reference technique, Radiofrequency Ablation (RFA). However, possible differences in lesion morphology between both techniques have been poorly studied. Simulations including electric, thermal and fluid physics were performed in a simplified model of the cardiac chamber which, in essence, consisted of a slab of myocardium with blood in motion on the top. Monopolar and bipolar catheter configurations were studied. Different blood velocities and catheter orientations were assayed. RFA was simulated assuming a conventional temperature-controlled approach. The PFA treatment was assumed to consist in a sequence of 20 biphasic bursts (100 µs duration). Simulations indicate that, for equivalent lesion depths, PFA lesions are wider, larger and more symmetrical than RFA lesions for both catheter configurations. RFA lesions display a great dependence on blood velocity while PFA lesions dependence is negligible on it. For the monopolar configuration, catheter angle with respect to the cardiac surface impacted both ablation techniques but in opposite sense. The orientation of the catheter with respect to blood flow direction only affected RFA lesions. In this study, substantial morphological differences between RFA and PFA lesions were predicted numerically. Negligible dependence of PFA on blood flow velocity and direction is a potential important advantage of this technique over RFA., This work was partially supported by project (PID2019-110120RBI00/AEI/10.13039/501100011033) from the Ministry of Science, Innovation and Universities and the State Research Agency of the Spanish government. Mario Gómez-Barea research is supported by a scholarship from the UPF. Tomás García-Sánchez acknowledges the funding received from the European Union’s Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie grant agreement No 892393 (TACAIRE). Antoni Ivorra gratefully acknowledges the financial support by ICREA under the ICREA Academia programme.