NEMA NU 4-2008 performance and MRI-compatibility study of the edgeless preclinical PET insert: ScintoTube

  • Gonzalez-Montoro, Andrea
  • Lopez-Berenguer, Fernando
  • Alamo. Jorge
  • Barberá, Julio
  • Berr, Stuart S.
  • Correcher, Carlos
  • Rispoli, Joseph V.
  • Sachs, Jennifer L.
  • Vidal, Luis F.
  • Williams, Mark B.
  • Benlloch, Jose M.
  • Gonzalez, Antonio J.
  • Freire-López-Fando, Marta
  • Alfonso Laguna, Carlos De|||0000-0002-2378-021X
  • Moliner Martínez, Laura
[EN] Objective. The goal of this work is to evaluate the performance of a preclinical Positron emission tomography (PET) system, named ScintoTube, which was constructed using a continuous (edgeless) LYSO:Ce scintillator. The PET compatibility tests with high-field magnetic resonance imaging (MRI) scanners are also reported. Approach. We constructed a preclinical PET system based on a single, continuous-annular LYSO:Ce scintillator, with inner and outer diameters of 64 mm and 80 mm, respectively, and an axial coverage of 96 mm. The system has 24 virtual detectors, being each one composed of a matrix of 9 × 9 silicon photomultipliers. A novel trigger topology was implemented to retrieve the entire light distribution (LD) profiles. This information was used to provide the 3D photon impact coordinates which included the depth of interaction (DOI) information. To evaluate the ScintoTube performance, the National Electrical Manufacturers Association NU 4-2008 protocol was followed. Dead time and quantification corrections were implemented in the reconstruction process. Moreover, since the system is intended to be used as an insert for high-field MRI scanners, it was evaluated when working under the influence of 7 T and 9.4 T MRI scanners by acquiring data with different MRI sequences. Results. This work successfully implements the edgeless concept for small animal PET insert and demonstrates that it is possible to retrieve homogeneous 3D photon impact positioning accuracy (in the 1 mm range) across the axial field of view by using an edgeless design. The reported results also validate the capabilities of the PET insert to work under the influence of high-field MRI. There is almost no influence of the MR in our PET insert regarding spatial and energy resolutions and photopeak position. All PET parameter deviations are within the ±5% range when compared to non-MRI case. Significance. Overall, the obtained results show that the designed ScintoTube is a promising system to be used as an insert with unique capabilities such as fully characterizing the entire LD profiles (3D positioning), suppressing edge effects in the transaxial and axial directions, and improving sensitivity., Andrea Gonzalez-Montoro is supported by the Spanish Ministry of Science and Innovation and the European Social Fund Project: RYC2021-031744-I.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Timing evaluation of a PET detector block based on semi-monolithic LYSO crystals

  • Cucarella-Melo, Neus
  • Barrio-Toala, John|||0000-0001-5181-2146
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • González Martínez, Antonio Javier
  • Lamprou, Efthymios
  • Valladares, Celia
[EN] Purpose Detectors for positron emission tomography (PET) typically use two types of scintillation crystals, pixelated or monolithic. A variant of these types of scintillators are the so-called semi-monolithic crystals. They consist of a monolithic crystal segmented in one direction in pieces called slabs. These scintillators have the potential to successfully combine the benefits of pixelated and monolithic configurations, providing good timing and spatial resolutions as well as the capacity to decode the depth of interaction (DOI) information. In this work, the timing performance of a detector based on semi-monolithic crystals was studied in depth. The energy response was also evaluated. Methods The semi-monolithic detector consists of 1 x 24 LYSO slabs of 25.4 x 12 x 0.95 mm(3) each. The bottom surface of the slabs is coupled to an array of 8 x 8 silicon photomultipliers (SiPMs) of 3 x 3 mm(2) active area, 50 mu m cell size and 3.2 mm pitch. The 64 output signals were independently readout by the TOFPET2 ASIC. In order to achieve the best coincidence time resolution (CTR), four different time walk corrections were tested. Additional work investigated the best method of combining the timestamps belonging to the same event. Results The resolvability of the slabs in the measured flood maps improves with the thickness of a light guide placed in between the scintillators and photosensors. The energy resolution does not change significantly with values as good as 13.7%. Regarding the CTR, values of 335.8, 363, 369.8, and 402.5 ps have been obtained for the whole detector for no light guide, 0.5, 1.0, and 1.5 mm thickness light guide cases, respectively. These values further improve to 276.1, 302.6, 305.6 and 336.2 ps, respectively, when energy-weighted averaging of timestamps is applied. Conclusions We have shown both an excellent timing resolution and good energy resolution for a PET detector based on semi-monolithic crystals. The use of light guides of different thicknesses does not significantly affect the energy resolution of the whole detector, but the timing capabilities slightly worsen with the increasing thickness of the light guide., European Research Council (ERC), Grant/Award Number: 695536; USA National Institutes of Health (NIH), Grant/Award Number: 1R01EB029450-01
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Evaluation of a PET Insert for Trimodal Imaging: A Step Towards PET/MRI-Guided Focused Ultrasound

  • Lopez-Berenguer, F.
  • Gonzalez-Montoro, Andrea
  • González-Tamarit, A.
  • Loignon-Houle, F.
  • Torres, I.
  • Sachs, J.L.
  • Rispoli, J.V
  • Tustison, N.
  • Berr, S. S.
  • Williams, M. B.
  • Andrés, D.
  • Guallart-Naval, T.
  • Alonso-Otamendi, Joseba
  • Sheybani, N. D.
  • Maslova, S.
  • Sherlock, T.
  • Thede, A.
  • Demir, Z.E.F.
  • Hoch, M.
  • He, J.
  • Freire-López-Fando, Marta
  • Jiménez-Serrano, Santiago|||0000-0003-2917-6053
  • Vidal San Sebastian, Luis Fernando|||0000-0001-8852-8560
  • Gil-Cervera, Marc
  • Rodríguez-García, Juan José
  • Alonso-Ramos, José Luis
  • Jimenez, Noe|||0000-0002-6539-670X
  • CARRIÓN GARCÍA, ALICIA|||0000-0002-0630-6065
  • Camarena Femenia, Francisco|||0000-0002-6713-1414
  • Algarín-Guisado, José Miguel|||0000-0003-2644-7225
  • Bosch-Esteve, Rubén|||0000-0002-7316-7819
  • Pallás-Lodeiro, Eduardo
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • González Martínez, Antonio Javier
[EN] Combining Positron Emission Tomography (PET) with Magnetic Resonance Imaging (MRI) and Focused Ultrasound (FUS) has emerged as a promising hybrid technique in the medical imaging field. Despite the potential benefits of using simultaneous PET and MRI acquisitions to monitor therapeutic effects and ensure precision and safety during FUS applications, no commercial or academic PET/MRI-guided FUS system (referred in the following as trimodal PET-MRI-FUS) exists. This work presents the design and evaluation of a preclinical PET insert for simultaneous operation with MRI and FUS systems. The proposed PET insert is based on monolithic LYSO crystals arranged in two octagonal rings that define inner and outer diameters of 72 mm and 114 mm, respectively, with an axial length of 67 mm. The system performance was evaluated according to the NEMA NU 4 2008 protocol. It achieves a uniform spatial resolution of ~0.9 mm along the whole PET Field of View (FOV) when enabling Depth of Interaction (DOI) capabilities. Moreover, we found a sensitivity peak of 3.8% for an energy window of 255-766 keV and a peak noise equivalent count rate peak of 80 kcps at an activity of 19 MBq. The compatibility of the PET insert with both low-and high-field MRI systems, as well as with custom and commercially available FUS devices, was studied using phantoms and in vivo experiments, respectively. The results of the trimodal tests have proven that the PET insert works simultaneously with both MRI and FUS systems., The authors would like to thank the Bruker team for their continued technical support, and Amadeo Ten-Esteve, the Biomedical Imaging Research Group at La Fe Hospital, for their support during data collection at La Fe Hospital. This work was supported by the Generalitat Valenciana under Grant No. CIPROM/2021/003 and funding from the University of Virginia Focused Ultrasound Cancer Immunotherapy Center and the FUS Foundation. Andrea Gonzalez-Montoro is supported by the Spanish Ministry of Science and Innovation and the European Social Fund project: RYC2021-031744-I. We acknowledge additional support from National Institutes of Health awards DP5OD031846 to N.D.S., T32CA009109 to Z.E.F.D. and T.S., and R01EB029450 to S.S.B. and A.J.G.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Impact on PET spatial resolution through positron range confinement in a high magnetic field across tissue-equivalent materials

  • Lopez-Berenguer, Fernando
  • Gonzalez-Montoro, Andrea
  • Berr, S. S.
  • Williams, M. B.
  • Freire-López-Fando, Marta
  • González Martínez, Antonio Javier
[EN] Objective. The spatial resolution of positron emission tomography (PET) imaging is intrinsically limited by the finite range of positrons before annihilation, an effect that becomes increasingly relevant for high-energy emitters and for preclinical studies targeting small structures. In integrated PET/ magnetic resonance imaging (MRI) systems, the magnetic field can modify positron trajectories, reducing their transverse spread and introducing anisotropic resolution effects. However, experimental evidence of the combined influence of positron energy and material density under high-field preclinical conditions remains limited. This work experimentally investigates the effect of a 9.4 T magnetic field on positron-range blurring for different radionuclides and tissue-equivalent phantoms in a preclinical PET/MRI environment. Approach. Experiments were performed using a high-resolution preclinical PET insert operating simultaneously inside a preclinical 9.4 T MRI system to study three positron-emitting radionuclides (18F, Zr-8(9) and 6(8)Ga). Capillary line sources were embedded in three tissue-equivalent phantoms with increasing density. Spatial resolution was quantified along the three spatial directions using the full width at half maximum and full width at tenth maximum (FWTM). Complementary measurements with a microDerenzo phantom were performed to assess rod resolvability. Main results. In low-density material, spatial resolution remains essentially unchanged by the magnetic field for all radionuclides. In contrast, in medium- and high-density materials, a marked transverse confinement is observed at 9.4 T for 6(8)Ga, with transverse FWTM values reduced by about 60% compared with 0 T, while the axial component remains largely unaffected. For Zr-8(9), transverse improvements of about 25% are observed. These trends are consistent with the microDerenzo results: for 6(8)Ga, no rod sector is resolvable at 0 T, whereas at 9.4 T rods close to 1.0 mm become resolvable. Significance. This study provides a systematic experimental assessment of positron-range confinement in high-field preclinical PET/MRI and demonstrates that the magnitude of magnetic-field-induced resolution improvements depends on both positron energy and material density., This work was supported by the Generalitat Valenciana under Grant No. CIPROM/2021/003 and Funding from the University of Virginia Focused Ultrasound Cancer Immunotherapy Center and the FUS Foundation. We acknowledge additional support from the National Institutes of Health award R01EB029450 to S.S.Berr and A.J.Gonzalez. A. Gonzalez-Montoro is supported by the Spanish Ministry of Science and Innovation and the European Social Fund project: RYC2021-031744-I. M. Freire is supported by Generalitat Valenciana through CIAPOS/2023/141 postdoctoral grant. The experimental work carried out at the University of Virginia was supported through the iMOVE mobility programme of the Spanish National Research Council (CSIC).