RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

NEMA Performance Evaluation of CareMiBrain dedicated brain PET and Comparison with the whole-body and dedicated brain PET systems

  • Moliner Martínez, Laura
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • González Martínez, Antonio Javier
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • CATRET MASCARELL, JUAN VICENTE
  • Ilisie, Victor
[EN] This article presents system performance studies of the CareMiBrain dedicated brain PET according to NEMA NU 2-2012 (for whole-body PETS) and NU 4-2008 (for preclinical PETs). This scanner is based on monolithic LYSO crystals coupled to silicon photomultipliers. The results obtained for both protocols are compared with current commercial whole body PETs and dedicated brain PETs found in the literature. Spatial resolution, sensitivity, NECR and scatter-fraction are characterized with NEMA standards, as well as an image quality study. A customized image quality phantom is proposed as NEMA phantoms do not fulfil the necessities of dedicated brain PETs. The full-width half maximum of the radial/tangential/ axial spatial resolution of CareMiBrain reconstructed with FBP at 10 and 100 mm from the system center were, respectively, 1.87/1.68/1.39 mm and 1.86/1.91/1.40 mm (NU 2-2012) and 1.58/1.45/1.40 mm and 1.64/1.66/1.44 mm (NU 4-2008). Peak NECR was 49 kcps@287 MBq with a scatter fraction of 48% using NU 2-2012 phantom. The sensitivity was 13.82 cps/kBq at the center of the FOV (NU 2-2012) and 10% (NU 4-2008). Contrast recovery coefficients for customizing image quality phantom were 0.73/0.78/1.14/1.01 for the 4.5/6/9/12 mm diameter rods. The performance characteristics of CareMiBrain are at the top of the current technologies for PET systems. Dedicated brain PET systems significantly improve spatial resolution and sensitivity, but present worse results in count rate measurements and scatter-fraction tests. As for the comparison of preclinical and clinical standards, the results obtained for solid and liquid sources were similar., This study was funded by the Spanish Ministry of Science, Innovation and University under grant RTC-2016-5186-1, a project co-financed by the European Union through the European Regional Development Fund (ERDF). CareMiBrain system was funding from the European Union's Horizon 2020 research and innovation programme under grant agreement No. 711323. Author Dr. Jose Maria Benlloch owns a small percentage of Oncovision S.A. The other authors declare no potential conflict of interest.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

QR-Factorization Algorithm for Computed Tomography (CT): Comparison With FDK and Conjugate Gradient (CG) Algorithms

  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • F Sánchez
  • Moliner Martínez, Laura
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • Soriano Asensi, Antonio
  • Sánchez Góez, Sebastián
[EN] Even though QR-factorization of the system matrix for tomographic devices has been already used for medical imaging, to date, no satisfactory solution has been found for solving large linear systems, such as those used in computed tomography (CT) (in the order of 106 equations). In CT, the Feldkamp, Davis, and Kress back projection algorithm (FDK) and iterative methods like conjugate gradient (CG) are the standard methods used for image reconstruction. As the image reconstruction problem can be modeled by a large linear system of equations, QR-factorization of the system matrix could be used to solve this system. Current advances in computer science enable the use of direct methods for solving such a large linear system. The QR-factorization is a numerically stable direct method for solving linear systems of equations, which is beginning to emerge as an alternative to traditional methods, bringing together the best from traditional methods. QR-factorization was chosen because the core of the algorithm, from the computational cost point of view, is precalculated and stored only once for a given CT system, and from then on, each image reconstruction only involves a backward substitution process and the product of a vector by a matrix. Image quality assessment was performed comparing contrast to noise ratio and noise power spectrum; performances regarding sharpness were evaluated by the reconstruction of small structures using data measured from a small animal 3-D CT. Comparisons of QR-factorization with FDK and CG methods show that QR-factorization is able to reconstruct more detailed images for a fixed voxel size., This work was supported by the Spanish Government under Grant TEC2016-79884-C2 and Grant RTC-2016-5186-1.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Contribución de congreso (ConferenceOutput)

Motion Correction of Multi-Frame PET Data

  • Cañizares, Gabriel
  • Espinós-Morató, Héctor
  • Hernández, Ángel
  • Álvarez-Gómez, Juan Manuel
  • Santos-Blasco, Joaquín
  • Moliner Martínez, Laura
  • González Martínez, Antonio Javier
  • F Sánchez
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
[EN] Motion correction algorithms are necessary in PET studies where the patient cannot remain totally static. Otherwise, artifacts appear in the final reconstructed image, degrading the spatial resolution or even misplacing the radiation distribution source. In this work we propose a new enhanced motion correction algorithm (EMAF) based on the multiple acquisition frames (MAF). This methodology can be applied to two PET system geometries (a ring and a multi-panel system). Using simulated data, the comparison of static and motion corrected profiles shows minimum differences keeping more than 99% of the events after the process, with a minimal inter-frame error (the maximum range of loss counts is around 0.6%). To measure the quality of the correction method, two error metrics as peak signal to noise ratio (PSNR) and intensity matching precision (IMP) are proposed. A Mini-Derenzo in a ring PET system reveals a remarkable improvement both for these two metrics (44.5 dB and 98.7%) when compared to the uncorrected images (41.1 dB and 56%, respectively). An equivalent analysis for a point-like source also shows a sizeable enhancement both in contrast and resolution (64 dB and 98% versus 62.5 dB and 17.9%, PSNR and IMP correspondingly). The proposed algorithm minimizes the image artifacts and its simplicity, independency of PET configuration system and rapid reconstruction and registration times, makes it a useful tool in preclinical PET studies., This work has been co-financed by the Spanish Government grants TEC2016-79884-C2 and RTC-2016-5186-1, and by the European Union through the European Regional Development Fund (ERDF)
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Contribución de congreso (ConferenceOutput)

Deep learning for MRI-based CT synthesis: a comparison of MRI sequences and neural network architectures

  • Larroza, Andrés
  • Álvarez-Gómez, Juan Manuel
  • Espinós-Morató, Héctor
  • Vergara-Díaz, Marina
  • Moliner Martínez, Laura
  • Oliver-Gil, Sandra|||0000-0001-8258-3972
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
[EN] Synthetic computed tomography (CT) images derived from magnetic resonance images (MRI) are of interest for radiotherapy planning and positron emission tomography (PET) attenuation correction. In recent years, deep learning implementations have demonstrated improvement over atlasbased and segmentation-based methods. Nevertheless, several open questions remain to be addressed, such as which is the best of MRI sequences and neural network architectures. In this work, we compared the performance of different combinations of two common MRI sequences (T1- and T2-weighted), and three state-of-the-art neural networks designed for medical image processing (Vnet, HighRes3dNet and ScaleNet). The experiments were conducted on brain datasets from a public database. Our results suggest that T1 images perform better than T2, but the results further improve when combining both sequences. The lowest mean average error over the entire head (MAE = 101.76 ± 10.4 HU) was achieved combining T1 and T2 scans with HighRes3dNet. All tested deep learning models achieved significantly lower MAE (p < 0.01) than a well-known atlas-based method., This work was supported by the Spanish Government grants TEC2016-79884-C2 and RTC-2016-5186-1, and by the European Union through the European Regional Development Fund (ERDF)
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Building blocks of a multi-layer PET with time sequence photon interaction discrimination and double Compton camera

  • Ilisie, V.
  • Giménez-Alventosa, Vicent|||0000-0003-1646-6094
  • Moliner Martínez, Laura
  • F Sánchez
  • González Martínez, Antonio Javier
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
[EN] Current PET detectors have a very low sensitivity, of the order of a few percent. One of the reasons is the fact that Compton interactions are rejected. If an event involves multiple Compton scattering and the total deposited energy lays within the photoelectric peak, then an energy-weighted centroid is the given output for the coordinates of the reconstructed interaction point. This introduces distortion in the final reconstructed image. The aim of our work is to prove that Compton events are a very rich source of additional information as one can improve the resolution of the detector and implicitly the final reconstructed image. This could be a real breakthrough for PET detector technology as one should be able to obtain better results with less patient radiation. Using a PET as a double Compton camera, by means of Compton cone matching i.e., Compton cones coming from the same event should be compatible, is applied to discard randoms, patient scattered events and also, to perform a correct matching among events with multiple coincidences. In order to fully benefit experimentally from Compton events using monolithic scintillators a multi-layer configuration is needed and a good time-of-flight resolution., This project has received funding from the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation programme (Grant Agreement No. 695536). This work was supported in part by the Spanish Government Grants TEC2016-79884-C2 and RTC-2016-5186-1.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Simulation Study for Designing a Dedicated Cardiac TOF-PET System

  • Oliver-Gil, Sandra|||0000-0001-8258-3972
  • Moliner Martínez, Laura
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • Ilisie, V.
[EN] The development of dedicated positron emission tomography scanners is an active area of research, especially aiming at the improvement of lesion detection and in support of cancer treatment and management. Recently, dedicated Positron Emission Tomography (PET) systems with different configurations for specific organs have been developed for improving detection effectiveness. Open geometries are always subject to distortion and artifacts in the reconstructed images. Therefore, the aim of this work is to determine the optimal geometry for a novel cardiac PET system that will be developed by our team, and determine the time resolution needed to achieve reasonable image quality for the chosen geometry. The proposed geometries consist of 36 modules. These modules are arranged in two sets of two plates, each one with different configurations. We performed Monte Carlo simulations with different TOF resolutions, in order to test the image quality improvement in each case. Our results show, as expected, that increasing TOF resolution reduces distortion and artifact effects. We can conclude that a TOF resolution of the order of 200 ps is needed to reduce the artifacts, to acceptable levels, generated in the simulated cardiac-PET open geometries., This project has been co-financed by the Spanish Government Grants TEC2016-79884-C2 and RTC-2016-5186-1, by the European Union through the European Regional Development Fund (ERDF) and by the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program (Grant Agreement No. 695536). The work of V.I. was supported by the Generalitat Valenciana APOSTD/2019/086 fellowship.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

High resolution and sensitivity gamma camera with active septa. A first Monte Carlo study

  • Ilisie, Victor
  • Moliner Martínez, Laura
  • Oliver-Gil, Sandra|||0000-0001-8258-3972
  • F Sánchez
  • González Martínez, Antonio Javier
  • Seimetz, Michael
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
[EN] Gamma cameras are of great interest due to their high potential in the field of Nuclear Medicine Imaging. They allow for an early diagnosis of reduced size tumors, and also for a wide variety of preclinical studies with the aim of designing more effective treatments against cancer. In this work we propose a significantly improved multi-pinhole collimator gamma camera and perform a first Monte Carlo analysis of its characteristics. Maintaining the configuration of a multi-pinhole collimator with a high degree of overlapping (thus with a high sensitivity), we add a new element, an active septa, that besides acting as a collimator, is able to measure the impact coordinates of the incident photon. This way one is able to unambiguously identify through which pinhole any gamma ray passes before being detected. The result is a high sensitivity and resolution multi-pinhole gamma camera with an arbitrarily large field of view. As a consequence, the final reconstructed image does not suffer from the undesired artifacts or truncation associated to the multiplexing phenomenon. In this study we focus on the development of a system able to visualize in 3D tumors, nodes and metastasis in real time in the operating room with very low dose. We also briefly analyse and propose a novel design for a Single Photon Emission Computed Tomography system., This project has received funding from the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation programme (Grant Agreement No. 695536). This work was supported in part by the Spanish Government Grants Generalitat Valenciana, APOSTD/2019/086 and TEC2016-79884-C2 and RTC-2016-5186-1.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

TOF studies for dedicated PET with open geometries

  • Moliner Martínez, Laura
  • González Martínez, Antonio Javier
  • Oliver-Gil, Sandra|||0000-0001-8258-3972
  • Giménez-Alventosa, Vicent|||0000-0003-1646-6094
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • F Sánchez
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • Ilisie, V.
  • Gonzalez, A.
  • Cañizares, G.
  • Lamprou, E.
  • Alamo, J.
[EN] Recently, two novel PET devices have been developed with open geometries, namely: breast and prostate-dedicated scanners. The breast-dedicated system comprises two detector rings of twelve modules with a field of view of 170 mm x 170 mm x 94 mm. Each module consists of a continuous trapezoidal LYSO crystal and a PSPMT. The system has the capability to vary the opening of the rings up to 60 mm in order to allow the insertion of a needle to perform a biopsy procedure. The prostate system has an open geometry consisting on two parallel plates separated 28 cm. One panel includes 18 detectors organized in a 6 x 3 matrix while the second one comprises 6 detectors organized in a 3 x 2 matrix. All detectors are formed by continuous LYSO crystals of 50 mm x 50 mm x15 mm, and a SiPM array of 12 x 12 individual photo-detectors. The system geometry is asymmetric maximizing the sensitivity of the system at the prostate location, located at about 2/3 in the abdomen-anus distance. The reconstructed images for PET scanners with open geometries present severe artifacts due to this peculiarity. These artifacts can be minimized using Time Of Flight information (TOF). In this work we present a TOF resolution study for open geometries. With this aim, the dedicated breast and prostate systems have been simulated using GATE (8.1 version) with different TOF resolutions in order to determine the image quality improvements that can be achieved with the existing TOF-dedicated electronics currently present in the market. The images have been reconstructed using the LMOS algorithm including TOF modeling in the calculation of the voxel-Line Of Response emission probabilities., This work was supported in part by the Spanish Government Grants TEC2016-79884-C2 and RTC-2016-5186-1 and by the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program (Grant Agreement No. 695536).