Shedding Light on Halos: Quantifying the Impact of the Diffractive Profile in Multifocal Intraocular Lenses

  • Martínez-Espert, Anabel
  • Garcia-Delpech, Salvador
  • Furlan, Walter D.
  • Monsoriu Serra, Juan Antonio|||0000-0003-3350-7951
[EN] Purpose: Halos are the most common source of discomfort after implantation of diffractive multifocal intraocular lenses (MIOLs). Although multiple causes are attributed to halos, there is no clear consensus about the specific role of each factor, especially the diffractive profile. This study aims to quantify the isolated effect of the diffractive profile on halo formation using numerical modeling. Methods: A simplified model eye was implemented in Zemax OpticStudio to simulate the optical performance of two different diffractive designs (sinusoidal and apodized kinoform) obtained from commercially available MIOLs. Simulations were performed for three pupil diameters under both monochromatic (green) and polychromatic illumination. To evaluate halo size, the half intensity diameter (HD) metric was used. Results: At the far focus, the kinoform MIOL exhibited a reduction in halo size as the pupil diameter increased. In contrast, the sinusoidal MIOL performed best at the intermediate focus for 6.0 mm, whereas at 3.0 mm and 4.5 mm the HD was nearly three times higher. In both cases, no significant difference was observed between monochromatic and polychromatic illumination. Conclusions: The observed differences in halo formation between the diffractive profiles studied can be primarily attributed to the way each design distributes optical energy across its diffraction orders. Translational Relevance: This study shows that diffractive profile of MIOLs significantly affects halo formation, aiding surgeons in interpreting patient complaints and evaluating lens designs critically. This approach also offers a useful tool for designers to minimize halos in new lenses., Supported by Generalitat Valenciana (CIPROM/2022/30) ; Ministerio de Ciencia e Innovacion (PID2022-142407NB-I00) . A. M.-E. acknowledges financial support from Universitat de Valencia (UV-INV-PREDOC21-1915492) .
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Silver mean sequence in extended depth of focus intraocular lenses: a comparative study of kinoform and stepwise designs

  • Martínez-Espert, Anabel
  • Muñoz-Pérez, Francisco M.
  • Furlan, Walter D.
  • Ferrando, Vicente|||0000-0002-7434-7109
  • Garmendía-Martínez, Adrián|||0000-0003-3443-5144
  • Monsoriu Serra, Juan Antonio|||0000-0003-3350-7951
[EN] In this work, we present two new multifocal intraocular lens (MIOL) designs, both based on the silver mean kinoform diffractive lens. We demonstrate that a single aperiodic diffractive profile can be used to create two different MIOLs: one with a kinoform structure and the other with a stepwise profile. Quantitative assessment of the designs was carried out using the through focus modulation transfer function and the area under the modulation transfer function for the prediction of their visual performance. Our results show that both designs exhibit nearly identical optical performance at the design wavelength (lambda = 550 nm), though their intrinsic longitudinal chromatic aberration differs significantly. Given that diffractive extended depth of focus (EDoF) intraocular lenses are prone to image degradation due to dysphotopic phenomena, we also compared the halos generated by these two designs and found notable differences in their behavior. Furthermore, under photopic conditions, the proposed lens designs demonstrated the potential to achieve visual acuity values of 0.2 logMAR or better across a vergence range from approximately 0 to 2 D. Finally, to qualitatively assess the behavior of the MIOLs, an objective experimental evaluation was conducted using an adaptive optics visual simulator in a model eye. Experimental results align with the quantitative assessment of the proposed designs., Generalitat Valenciana (CIPROM/2022/30) ; Ministerio de Ciencia e Innovacion (PID2022-142407NB-I00) ; Universitat de Valencia (UV-INV-PREDOC21-1915492).
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Visual performance of a new trifocal intraocular lens design evaluated with a clinical adaptative optics visual simulator

  • Martínez-Espert, Anabel
  • Montagud-Martínez, Diego
  • Garcia-Delpech, Salvador
  • Furlan, Walter D.
  • Ferrando, Vicente|||0000-0002-7434-7109
  • Monsoriu Serra, Juan Antonio|||0000-0003-3350-7951
[EN] Objective: The digital transformation of daily routines has increased visual demands, especially at intermediate and near distances. The Devil multifocal intraocular lens (MIOL) design was developed to enhance intermediate visual performance. This design is a novel trifocal intraocular lens inspired by the fractal structure known as the 'Devil's staircase'. The aim of this research is to evaluate the visual performance of the Devil MIOL design in real patients. Methods: The visual acuity defocus curve was obtained from 25 eyes of healthy volunteers using the Visual Adaptive Optics Simulator (VAO, Voptica SL, Murcia, Spain). Additionally, images of optotypes simulating those seen through the Devil MIOL design were captured to qualitatively illustrate optotype perception at different vergence values. Results: Mean visual acuity values of the evaluation were 0.03 logMAR at 0.00 D vergence, 0.10 logMAR at +1.50 D vergence, and 0.11 logMAR at +3.00 D vergence, corresponding to far (optical infinity), intermediate (66.7 cm), and near foci (33.3 cm), respectively. The experimental outcomes were indeed slightly better than numerical results obtained previously in a model eye. Conclusions: The Devil MIOL design provides satisfactory visual acuity across three primary foci and produces extended depth of focus between the intermediate and near foci., This work was supported by Ministerio de Ciencia e Innovacion (PID2022-142407NB-I00); and Generalitat Valenciana (CIPROM/2022/30). A. M.-E. acknowledges financial support from Universitat de Valencia programa Atracciode Talent 2021 (UV-INV-PREDOC21-1915492). D. M.-M. also acknowledges the Margarita Salas grant from the Ministerio de Universidades, Spain, funded by the European Union-Next Generation EU.