Durable 3D murine ex vivo retina glaucoma models for optical coherence tomography
[EN] Durable and standardized phantoms with optical properties similar to native healthy
and disease-like biological tissues are essential tools for the development, performance testing,
calibration and comparison of label-free high-resolution optical coherence tomography (HR-OCT)
systems. Available phantoms are based on artificial materials and reflect thus only partially ocular
properties. To address this limitation, we have performed investigations on the establishment
of durable tissue phantoms from ex vivo mouse retina for enhanced reproduction of in vivo
structure and complexity. In a proof-of-concept study, we explored the establishment of durable
3D models from dissected mouse eyes that reproduce the properties of normal retina structures
and tissue with glaucoma-like layer thickness alterations. We explored different sectioning
and preparation procedures for embedding normal and N-methyl-D-aspartate (NMDA)-treated
mouse retina in transparent gel matrices and epoxy resins, to generate durable three-dimensional
tissue models. Sample quality and reproducibility were quantified by thickness determination
of the generated layered structures utilizing computer-assisted segmentation of OCT B-scans
that were acquired with a commercial HR-OCT system at a central wavelength of 905 nm and
analyzed with custom build software. Our results show that the generated 3D models feature
thin biological layers close to current OCT resolution limits and glaucoma-like tissue alterations
that are suitable for reliable HR-OCT performance characterization. The comparison of data
from resin-embedded tissue with native murine retina in gels demonstrates that by utilization of
appropriate preparation protocols, highly stable samples with layered structures equivalent to
native tissues can be fabricated. The experimental data demonstrate our concept as a promising
approach toward the fabrication of durable biological 3D models suitable for high-resolution
OCT system performance characterization supporting the development of optimized instruments
for ophthalmology applications., Horizon 2020 Project GALAHAD (732613); Horizon Europe Project SEQUOIA (101070062); Open
Access Publication Fund of the University of Münster