Resin-embedded murine ex vivo retina as durable models for optical coherence tomography
[EN]
High-resolution optical coherence tomography (HR-OCT) systems require standardized, biologically relevant phantoms to enable reproducible performance evaluation and benchmarking. In this study, we present an optimized method for preparing durable 3D retina phantoms from resin-embedded ex vivo mouse eyes. Mechanical polishing and cutting techniques were applied to improve sample geometry, enabling better positioning of retinal tissue relative to the resin¿air interface. The modified samples were characterized by using a commercial HR-OCT system operating at 905¿nm. Quantitative analysis of B-scan images revealed enhanced signal contrast, particularly in the upper retinal layers, due to reduced scattering. This approach improves the anatomical fidelity and imaging performance of retina phantoms, offering a promising tool for calibration and comparative assessment of advanced HR-OCT systems., Funding by the European Union through the programs Horizon 2020 (Project GALAHAD, no.: 732613) and Horizon