DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Intraocular lens alignment from an en face optical coherence tomography image Purkinje-like method

  • Sun, M.
  • Castro, Alberto de
  • Ortiz, Sergio
  • Pérez Merino, Pablo
  • Birkenfeld, Judith S.
  • Marcos, Susana
9 pags.; 6 figs.; 2 tabs.; . Published by SPIE under a Creative Commons Attribution 3.0 Unported License., Measurement of intraocular lens (IOL) alignment implanted in patients in cataract surgery is important to understand their optical performance. We present a method to estimate tilt and decentration of IOLs based on optical coherence tomography (OCT) images. En face OCT images show Purkinje-like images that correspond to the specular reflections from the corneal and IOL surfaces. Unlike in standard Purkinjeimaging, the tomographic nature of OCT allows unequivocal association of the reflection with the corresponding surface. The locations of the Purkinje-like images are linear combinations of IOL tilt, IOL decentration, and eye rotation. The weighting coefficients depend on the individual anterior segment geometry, obtained from the same OCT datasets. The methodology was demonstrated on an artificial model eye with set amounts of lens tilt and decentration and five pseudophakic eyes. Measured tilt and decentration in the artificial eye differed by 3.7% and 0.9%, respectively, from nominal values. In patients, average IOL tilt and decentration from Purkinje were 3.30 ± 4.68 deg and 0.16 ± 0.16 mm, respectively, and differed on average by 0.5 deg and 0.09 mm, respectively, from direct measurements on distortion-corrected OCT images. Purkinje-based methodology from anterior segment en face OCT imaging provided, therefore, reliable measurements of IOL tilt and decentration. © The Authors., This study has been funded by Spanish Government Grant No. FIS2011-25637 and European Research Council ERC-2011 AdG-294099 “Bio-inspired optical correction of presbyopia” to S. Marcos, and Marie Curie Actions-Networks of Initial Training (ITN) FP7-PEOPLE-2010-ITN 264605 (OpAL-Optical and adaptational limits of vision)., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Short-term neural adaptation to simultaneous bifocal images

  • Radhakrishnan, Aiswaryah
  • Dorronsoro, Carlos
  • Sawides, L.
  • Marcos, Susana
Simultaneous vision is an increasingly used solution for the correction of presbyopia (the age-related loss of ability to focus near images). Simultaneous Vision corrections, normally delivered in the form of contact or intraocular lenses, project on the patient's retina a focused image for near vision superimposed with a degraded image for far vision, or a focused image for far vision superimposed with the defocused image of the near scene. It is expected that patients with these corrections are able to adapt to the complex Simultaneous Vision retinal images, although the mechanisms or the extent to which this happens is not known. We studied the neural adaptation to simultaneous vision by studying changes in the Natural Perceived Focus and in the Perceptual Score of image quality in subjects after exposure to Simultaneous Vision. We show that Natural Perceived Focus shifts after a brief period of adaptation to a Simultaneous Vision blur, similar to adaptation to Pure Defocus. This shift strongly correlates with the magnitude and proportion of defocus in the adapting image. The magnitude of defocus affects perceived quality of Simultaneous Vision images, with 0.5 D defocus scored lowest and beyond 1.5 D scored >sharp>. Adaptation to Simultaneous Vision shifts the Perceptual Score of these images towards higher rankings. Larger improvements occurred when testing simultaneous images with the same magnitude of defocus as the adapting images, indicating that wearing a particular bifocal correction improves the perception of images provided by that correction. © 2014 Radhakrishnan et al., The research leading to these results has received funding from the European Research Council under the European Union's Seventh Framework Programme (FP7/2007-2013)/ERC Grant Agreement n° 294099, Spanish Government grant FIS2011-264605 and 7th Framework Programme of the European Community through the Marie Curie Initial Training Network OpAL (OpAL is an Initial Training Network funded by the European Commission under the Seventh Framework Programme (PITN-GA-2010-264605)), Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Single neural code for blur in subjects with different interocular optical blur orientation

  • Radhakrishnan, Aiswaryah
  • Sawides, L.
  • Dorronsoro, Carlos
  • Peli, E.
  • Marcos, Susana
13 págs.; 5 figs., © 2015 ARVO. The ability of the visual system to compensate for differences in blur orientation between eyes is not well understood. We measured the orientation of the internal blur code in both eyes of the same subject monocularly by presenting pairs of images blurred with real ocular point spread functions (PSFs) of similar blur magnitude but varying in orientations. Subjects assigned a level of confidence to their selection of the best perceived image in each pair. Using a classification-images–inspired paradigm and applying a reverse correlation technique, a classification map was obtained from the weighted averages of the PSFs, representing the internal blur code. Positive and negative neural PSFs were obtained from the classification map, representing the neural blur for best and worse perceived blur, respectively. The neural PSF was found to be highly correlated in both eyes, even for eyes with different ocular PSF orientations (rPos = 0.95; rNeg = 0.99; p < 0.001). We found that in subjects with similar and with different ocular PSF orientations between eyes, the orientation of the positive neural PSF was closer to the orientation of the ocular PSF of the eye with the better optical quality (average difference was ∼10°), while the orientation of the positive and negative neural PSFs tended to be orthogonal. These results suggest a single internal code for blur with orientation driven by the orientation of the optical blur of the eye with better optical quality., The authors acknowledge the funding received from the European Research Council under the European Union’s Seventh Framework Programme (FP7/2007- 2013) / ERC Grant Agreement n8 294099 (SM), Spanish Government grant FIS2011-264605 (SM), the 7th Framework Programme of the European Community through the Marie Curie Initial Training Network OpAL (OpAL is an Initial Training Network funded by the European Commission under the Seventh Framework Programme (PITN-GA-2010- 264605, AR) and USA NIH grant R01EY05957 (EP)., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Differences in visual quality with orientation of a rotationally asymmetric bifocal intraocular lens design

  • Radhakrishnan, Aiswaryah
  • Dorronsoro, Carlos
  • Marcos, Susana
12 págs.; 13 figs., Purpose To evaluate visual and perceptual performance for different orientations of a rotationally asymmetric bifocal intraocular lens (IOL) (M-Plus) simulated optically using a simultaneous vision simulator. Setting Instituto de Optica, Madrid, Spain. Design Prospective observational study. Methods Perceptual quality and decimal high-contrast visual acuity (HCVA) was measured under cycloplegia for 8 orientations of the asymmetric bifocal IOL phase pattern at far, intermediate, and near distances simulated with a simultaneous vision simulator using face images and tumbling E targets. The preferred orientation at each distance was calculated as the centroid of the data for 8 orientations. The visual Strehl value was calculated using the subjects' ocular aberrations and multifocal pattern at each orientation. Optical predictions were obtained by implementing a differential visual Strehl values–based ideal observer model. Results The study comprised 20 subjects (aged 21 to 62 years). Horizontal orientation (near segment at 0 or 180 degrees ± 45 [SD]) was preferred by 14 subjects and by 13 subjects at far and near distances, respectively; 8 subjects showed strong orientation preferences. The mean difference in preferred orientation between far and near was 27 ± 22 degrees. No significant differences in HCVA were observed. Optical predictions correlated strongly and significantly with measurements (far r = 0.71, near r = 0.62; P < .0001). The mean difference between measurement and simulation in the preferred orientation was 28 ± 29 degrees at far and 36 ± 28 degrees at near. Conclusions The perception varied for different orientations of an asymmetric bifocal IOL design tested using a simultaneous vision simulator. Optimum orientation was driven by interactions of the design with the eye's optical aberrations. Financial Disclosure None of the authors has a financial or proprietary interest in any material or method mentioned. Q 2016 ASCRS and ESCRS, Funding provided by the European Research Council under the European Union’s Seventh Framework Programme (FP7/2007- 2013)/ERC Grant Agreement number 294099 (Dr. Marcos), Spanish Government grant FIS2011-264605 and FIS2014-56643-R (Dr. Marcos), the 7th Framework Programme of the European Community through the Marie Curie Initial Training Network OpAL (OpAL is an Initial Training Network funded by the European Commission under the Seventh Framework Programme), (PITN-GA-2010- 264605, Dr. Radhakrishnan), and Oculentis GmBH, Germany., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Full 3-D OCT-based pseudophakic custom computer eye model

  • Sun, M.
  • Pérez Merino, Pablo
  • Martínez-Enríquez, Eduardo
  • Velasco Ocaña, Miriam
  • Marcos, Susana
15 págs.; 8 figs.; 4 tabs. ; OCIS codes: (110.4500) Optical coherence tomography; (200.4560) Optical data processing; (220.1010) Aberrations (global); (330.7326) Visual optics, modeling., We compared measured wave aberrations in pseudophakic eyes implanted with aspheric intraocular lenses (IOLs) with simulated aberrations from numerical ray tracing on customized computer eye models, built using quantitative 3-D OCT-based patient-specific ocular geometry. Experimental and simulated aberrations show high correlation (R = 0.93; p<0.0001) and similarity (RMS for high order aberrations discrepancies within 23.58%). This study shows that full OCT-based pseudophakic custom computer eye models allow understanding the relative contribution of optical geometrical and surgically-related factors to image quality, and are an excellent tool for characterizing and improving cataract surgery. ©2016 Optical Society of America, This research has received funding from the European Research Council under the European Union’s Seventh Framework Program (FP/2007-2013) / ERC Grant Agreement. [ERC-2011- AdC 294099], from Marie Curie Actions-Networks of Initial Training (ITN) FP7-PEOPLE-2010-ITN 264605 (OpAL-Optical and adaptational limits of vision), and from Spanish Government grants FIS2011-25637 and FIS2014-56643-R to SM. The authors indicate the following financial disclosure(s): Spanish patent: Procedure to calibrate and correct the scan distortion of an optical coherence tomography system, P201130685 (Sergio Ortiz, Susana Marcos, Damian Siedlecki, and Carlos Dorronosoro)., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Optical coherence tomography based estimates of crystalline lens volume, equatorial diameter, and plane position

  • Martínez-Enríquez, Eduardo
  • Sun, M.
  • Velasco Ocaña, Miriam
  • Birkenfeld, Judith S.
  • Pérez Merino, Pablo
  • Marcos, Susana
11 págs.; 8 figs. ; Open Access funded by Creative Commons Atribution Licence 4.0, PURPOSE. Measurement of crystalline lens geometry in vivo is critical to optimize performance of state-of-the-art cataract surgery. We used custom-developed quantitative anterior segment optical coherence tomography (OCT) and developed dedicated algorithms to estimate lens volume (VOL), equatorial diameter (DIA), and equatorial plane position (EPP). METHODS. The method was validated ex vivo in 27 human donor (19-71 years of age) lenses, which were imaged in three-dimensions by OCT. In vivo conditions were simulated assuming that only the information within a given pupil size (PS) was available. A parametric model was used to estimate the whole lens shape from PS-limited data. The accuracy of the estimated lens VOL, DIA, and EPP was evaluated by comparing estimates from the whole lens data and PS-limited data ex vivo. The method was demonstrated in vivo using 2 young eyes during accommodation and 2 cataract eyes. RESULTS. Crystalline lens VOL was estimated within 96% accuracy (average estimation error across lenses ± standard deviation: 9.30 ± 7.49 mm). Average estimation errors in EPP were below 40 ± 32 lm, and below 0.26 ± 0.22 mm in DIA. Changes in lens VOL with accommodation were not statistically significant (2-way ANOVA, P = 0.35). In young eyes, DIA decreased and EPP increased statistically significantly with accommodation (P < 0.001) by 0.14 mm and 0.13 mm, respectively, on average across subjects. In cataract eyes, VOL = 205.5 mm3, DIA = 9.57 mm, and EPP = 2.15 mm on average. CONCLUSIONS. Quantitative OCT with dedicated image processing algorithms allows estimation of human crystalline lens volume, diameter, and equatorial lens position, as validated from ex vivo measurements, where entire lens images are available., Supported by Ministerio de Educación y Ciencia, Spain, grants FIS2011-25637 and FIS2014-56643, European Research Council grant ERC-2011-AdG-294099 (SM), and Marie Curie ITN grant FP7- PEOPLE-2010-ITN 264605 (MS)., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Perceived image quality with simulated segmented bifocal corrections

  • Dorronsoro, Carlos
  • Radhakrishnan, Aiswaryah
  • Gracia, Pablo de
  • Sawides, L.
  • Marcos, Susana
12 págs.; 8 figs.; OCIS codes: (330.0330) Vision, color, and visual optics; (330.4300) Vision system - noninvasive assessment; (330.4460) Ophthalmic optics and devices; (330.4595) Optical effects on vision; (330.7327); Visual optics, ophthalmic instrumentation., Bifocal contact or intraocular lenses use the principle of simultaneous vision to correct for presbyopia. A modified two-channel simultaneous vision simulator provided with an amplitude transmission spatial light modulator was used to optically simulate 14 segmented bifocal patterns (+ 3 diopters addition) with different far/near pupillary distributions of equal energy. Five subjects with paralyzed accommodation evaluated image quality and subjective preference through the segmented bifocal corrections. There are strong and systematic perceptual differences across the patterns, subjects and observation distances: 48% of the conditions evaluated were significantly preferred or rejected. Optical simulations (in terms of through-focus Strehl ratio from Hartmann-Shack aberrometry) accurately predicted the pattern producing the highest perceived quality in 4 out of 5 patients, both for far and near vision. These perceptual differences found arise primarily from optical grounds, but have an important neural component. © 2016 Optical Society of America, European Research Council (ERC) (ERC-AdG-294099) and Spanish Government. Plan Nacional de I+D+i (FIS2011-264605, FIS2014-56643-R) to SM; EU FP7 Marie Curie Initial Training Network OpAL (PITN-GA-2010-264605) to AR; CSIC JAE-Pre Program to PdG; Spanish Government. MEyC FPI Predoctoral Fellowship to LS., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Portable simultaneous vision device to simulate multifocal corrections

  • Dorronsoro, Carlos
  • Radhakrishnan, Aiswaryah
  • Alonso-Sanz, J. R.
  • Pascual, Daniel
  • Velasco Ocaña, Miriam
  • Pérez Merino, Pablo
  • Marcos, Susana
7 págs.; 6 figs. ; OCIS codes: (330.0330) Vision, color, and visual optics; (330.4300) Vision system - noninvasive assessment; (330.4460) Ophthalmic optics and devices; (330.4595) Optical effects on vision; (330.7327) Visual optics, ophthalmic instrumentation., Multifocal lenses are increasingly used solutions for presbyopia, the age-related loss of crystalline lens focus ability. These lenses work by the principle of simultaneous vision, superimposing focused and defocused images on the retina. Providing the experience of simultaneous vision to a patient before permanent implantation of a multifocal lens is a recognized unmet need to increase the patient’s confidence and optimize the lens selection. We developed a hand-held, see-through multifocal vision simulator based on temporal multiplexing of a tunable lens. The device was calibrated and validated using focimetry and Hartmann–Shack aberrometry revealing high reproducibility of the through-focus multifocal energy distribution and high optical quality. We measured visual acuity and perceptual quality on nine cyclopeged patients with three monofocal, two bifocal, and two trifocal corrections with different far/intermediate/ near energy distributions simulated using the device. Visual performance and perceptual quality with multifocal corrections varied across patients, although they were more uniform across distances than monofocal corrections. Among the bifocal and trifocal designs, a trifocal with more energy at far was the most frequently identified as providing better quality. The simultaneous vision simulator proved a promising compact tool to study visual performance with multifocal corrections and to select the lens design best suited for each patient, alternative to costly and bulky adaptive optics based devices. © 2016 Optical Society of America, European Research Council (ERC) (ERC-2011- AdG-294099); Ministerio de Economía y Competitividad (MINECO) (FIS2011-264605, FIS2014-56643-R); Seventh Framework Programme (FP7) (PITN-GA-2010-264605)., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Role of parafovea in blur perception

  • Venkataraman, A. P.
  • Radhakrishnan, Aiswaryah
  • Dorronsoro, Carlos
  • Lundström, L.
  • Marcos, Susana
7 pags., 7 figs. -- Open Access funded by Creative Commons Atribution Licence 4.0, The blur experienced by our visual system is not uniform across the visual field. Additionally, lens designs with variable power profile such as contact lenses used in presbyopia correction and to control myopia progression create variable blur from the fovea to the periphery. The perceptual changes associated with varying blur profile across the visual field are unclear. We therefore measured the perceived neutral focus with images of different angular subtense (from 4° to 20°) and found that the amount of blur, for which focus is perceived as neutral, increases when the stimulus was extended to cover the parafovea. We also studied the changes in central perceived neutral focus after adaptation to images with similar magnitude of optical blur across the image or varying blur from center to the periphery. Altering the blur in the periphery had little or no effect on the shift of perceived neutral focus following adaptation to normal/blurred central images. These perceptual outcomes should be considered while designing bifocal optical solutions for myopia or presbyopia., This work was supported by the European Research Council under the 7th Framework Programme through the Marie Curie Initial Training Network OpAL (PITNGA- 2010-264605), the Swedish Research Council (621-2011-4094), European Union’s Seventh Framework Programme (FP7/2007-2013), ERC Grant Agreement (ERC-2011- AdC 294099), Spanish Government grants FIS2011-25637 and FIS2014-56643., Peer Reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Contribución de congreso (ConferenceOutput)

Three dimensional biometry and alignment in eyes implanted with Accommodative IOLs as a function of accommodative demand

  • Marcos, Susana
  • Ortiz, Sergio
  • Pérez Merino, Pablo
  • Velasco Ocaña, Miriam
  • Sun, M.
  • Birkenfeld, Judith S.
  • Durán, Sonia
  • Jiménez-Alfaro, Ignacio
ARVO Annual Meeting 2013, Washington State Convention Center, May 5-9, 2013, Spanish Government Grant FIS2011-25637; European Research Council ERC2011 AdG-294099; EU Marie Curie FP7-PEOPLE-2010-ITN 264605; CSIC JAE-Program; Unidad Asociada IO-CSIC/FJD