RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Radiofrequency Ablation Using a Novel Insulated-Tip Ablation Catheter Can Create Uniform Lesions Comparable in Size to Conventional Irrigated Ablation Catheters While Using a Fraction of the Energy and Irrigation

  • Aryana, Arash
  • Irastorza, Ramiro M.
  • Cohen, Richard J.
  • Kraus, Jeffrey
  • Haghighi-Mood, Ali
  • Reddy, Vivek Y.
  • D Avila, Andre
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Introduction: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), similar to 90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated-tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energy into the targeted tissue while utilizing reduced power and irrigation. Methods: This study evaluated the outcomes of RFA using SMT versus an RFC in silico, ex vivo, and in vivo. Radiofrequency applications were delivered over porcine myocardium (ex vivo) and porcine thigh muscle preparations superfused with heparinized blood (in vivo). Altogether, 274 radiofrequency applications were delivered using SMT (4-15 W, 2 or 20 ml/min) and 74 applications using RFC (30 W, 30 ml/min). Results: RFA using SMT proved capable of directing 66.8% of the radiofrequency energy into the targeted tissue. Accordingly, low power-low irrigation RFA using SMT (8-12 W, 2 ml/min) yielded lesion sizes comparable with RFC, whereas high power-high irrigation (15 W, 20 ml/min) RFA with SMT yielded lesions larger than RFC (p < .05). Although SMT was associated with greater impedance drops ex vivo and in vivo, ablation using RFC was associated with increased charring/steam pop/ tissue cavitation (p < .05). Lastly, lesions created with SMT were more homogeneous than RFC (p < .001). Conclusion: Low power-low irrigation (8-12 W, 2 ml/min) RFA using the novel SMT ablation catheter can create more uniform, but comparable-sized lesions as RFC with reduced charring/steam pop/tissue cavitation. High power-high irrigation (15 W, 20 ml/min) RFA with SMT yields lesions larger than RFC., Sirona Medical Technologies, Inc; The Proyecto UNAJ Investiga 2017, Grant/Award Number: 80020170100019UJ; The Spanish Ministerio de Ciencia, Innovacion y Universidades/Agencia Estatal de Investigacion (MCIN/AEI/10.13039/501100011033), Grant/Award Number: RTI2018-094357B-C21
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Effect of the relative position of electrode and stellate ganglion during thermal radiofrequency ablation: a simulation study

  • Irastorza, Ramiro M.
  • Bovaira, Maite
  • García-Vitoria, Carles
  • Muñoz, Víctor
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Purpose Stellate ganglion (SG) block by thermal radiofrequency ablation (RFA) is frequently conducted as a therapeutic intervention for sympathetic-maintained and neuropathic pain syndromes. RFA's partial lack of effectiveness could be partly due to the ablation zone (AZ) not completely covering the SG section and therefore preventing the 'cutting' of the afferent pathways. Our objective was to build a theoretical model to conduct computer simulations to assess the effect of the electrode position relative to the SG. Methods A three-dimensional model was built including the SG and adjacent tissues (vertebrae C7-T1-T2, trachea, carotid artery and vertebral artery). RFA (90-s, 80 degrees C) was simulated considering a 22 G-5 mm electrode. The AZ was computed using the 50 degrees C isotherm. Results An electrode displacement of 2 mm in any direction from the optimal position (centered on the SG) meant that the AZ did not fully cover the SG section. Likewise, SG size considerably affected the RFA effectiveness since the AZ fully covered the section of small but not large SGs. Conclusions The findings suggest that the currently used SG RFA settings (i.e., 22 G-5 mm electrode, 90-s, 80 degrees C) may not be appropriate due to their inability to achieve an AZ that fully covers the SG cross section under certain circumstances, such as a large SG and non-optimal positioning of the RF electrode with respect to the SG center., This work was supported by the Spanish Ministerio de Ciencia, Innovacion y Universidades under the 'Programa Estatal de I+D+i orientada a los retos de la Sociedad', Grant No 'RTI2018-094357-B-C21', the Universidad Nacional Arturo Jauretche - Argentina 'UNAJ Investiga' [80020170100019UJ], and by the 'Agencia Nacional de Promocion Cientifica y Tecnologica de Argentina' [PICT-2016-2303].
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publication Artículo científico (JournalArticle)

Differences in applied electrical power between full thorax models and limited domain models for RF cardiac ablation

  • Irastorza, Ramiro M.
  • González-Suárez, Ana|||0000-0002-1813-4176
  • Pérez, Juan J|||0000-0001-8486-8699
  • Berjano, Enrique|||0000-0002-3247-2665
[EN] Purpose:Most modeling studies on radiofrequency cardiac ablation (RFCA) are based on limited-domain models, which means the computational domain is restricted to a few centimeters of myocardium and blood around the active electrode. When mimicking constant power RFCA procedures (e.g., atrial fibrillation ablation) it is important to know how much power is absorbed around the active electrode and how much in the rest of the tissues before reaching the dispersive electrode. Methods:3D thorax full models were built by progressively incorporating different organs and tissues with simplified geometries (cardiac chamber, cardiac wall, subcutaneous tissue and skin, spine, lungs and aorta). Other 2D limited-domain models were also built based on fragments of myocardium and blood. The electrical problem was solved for each model to estimate the spatial power distribution around the active electrode. Results:From 79 to 82% of the power was absorbed in a 4 cm-radius sphere around the active electrode in the full thorax model at active electrode insertion depths of between 0.5 and 2.5 mm, while the impedance values ranged from 104 to 118 omega, which were consistent with those found (from 83 to 103 omega) in a 4 cm radius cylindrical limited domain model. Conclusion:The applied power in limited-domain RFCA models is approximately 80% of that applied in full thorax models, which is equivalent to the power programed in a clinical setting., This work was supported by the Spanish Ministerio de Economia, Industria y Competitividad under `Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad', Grant No [RTI2018094357-B-C21], by the Universidad Nacional Arturo Jauretche Argentina 'UNAJ Investiga' [80020170100019UJ], and by `Agencia Nacional de Promocion Cientifica y Tecnologica de Argentina' [PICT2016-2303].