Skip to main content

DESARROLLO DE UNA HERRAMIENTA COMPUTACIONAL DE INTERACCION FLUIDO-SOLIDO EN EL AMBITO CARDIOVASCULAR. APLICACION AL DISEÑO DE UN NUEVO PROTOTIPO DE FILTRO ANTITROMBO

DPI2010-20746-C03-01

•
Funding agency name Ministerio de Ciencia e Innovación
Funding agency acronym MICINN
Program Programa Nacional de Proyectos de Investigación Fundamental
Subprogram Subprograma de Proyectos de Investigación Fundamental no Orientada
Call Proyectos I+D
Call year 2010
State Plan 2008-2011 Plan Nacional de Investigación Científica, Desarrollo e Innovación Tecnológica
Management unit Subdirección General de Proyectos de Investigación
Beneficiary centre UNIVERSIDAD DE ZARAGOZA

Resultados relacionados

Resultados totales (Incluyendo duplicados): 5
Encontrada(s) 1 página(s)

Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Numerical and experimental study of the fluid flow through a medical device

  • Nicolás, M.
  • Palero, V.R.
  • Peña, E.
  • Arroyo, M.P.
  • Martínez, M.A.
  • Malvè, M.
The purpose of this paper is to verify a commercial software based fluid–structure interaction scheme for the inferior vena cava. Vena cava deep thrombosis (TVP) is a potentially deathly disease consequent to pulmonary thromboembolism (TEP). TEP consist in the obstruction of the pulmonary artery due to a blood clot traveling in the cardiovascular system and is treated with anticoagulants and inferior vena cava filters. Flow fields along the vena cava and an antithrombus filter were studied and compared with a Particle Image Velocimetry (PIV) based model to validate the numerical model. The results show that the fluid–structure interaction (FSI) models are valid and can be used to study the deformations in the inferior vena cava wall using patient-specific geometries.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

In vitro comparison of Günther Tulip and Celect filters. Testing filtering efficiency and pressure drop

  • Nicolas, M.
  • Malvé, M.
  • Peña, E.
  • Martínez, M.A.
  • Leask, R.
In this study, the trapping ability of the Günther Tulip and Celect inferior vena cava filters was evaluated. Thrombus capture rates of the filters were tested in vitro in horizontal position with thrombus diameters of 3 and 6. mm and tube diameter of 19. mm. The filters were tested in centered and tilted positions. Sets of 30 clots were injected into the model and the same process was repeated 20 times for each different condition simulated. Pressure drop experienced along the system was also measured and the percentage of clots captured was recorded. The Günther Tulip filter showed superiority in all cases, trapping almost 100% of 6. mm clots both in an eccentric and tilted position and trapping 81.7% of the 3. mm clots in a centered position and 69.3% in a maximum tilted position. The efficiency of all filters tested decreased as the size of the embolus decreased and as the filter was tilted. The injection of 6 clots raised the pressure drop to 4.1. mmHg, which is a reasonable value that does not cause the obstruction of blood flow through the system.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Influence of a commercial antithrombotic filter on the caval blood flow during neutra and valsalva maneuver

  • Nicolas, M.
  • Lucea, B.
  • Laborda García, Alicia
  • Peña, E.
  • Gregorio, M.A. de
  • Martinez, M.A.
  • Malve, M.
Anticoagulants are the treatment of choice for pulmonary embolism. When these fail or are contraindicated, vena cava filters are effective devices for preventing clots from the legs from migrating to the lung. Many uncertainties exist when a filter is inserted, especially during physiological activity such as normal breathing and the Valsalva maneuver. These activities are often connected with filter migration and vena cava damage due to the various related vein geometrical configurations. In this work, we analyzed the response of the vena cava during normal breathing and Valsalva maneuver, for a healthy vena cava and after insertion of a commercial Günther-Tulip® filter. Validated computational fluid dynamics (CFD) and patient specific data are used for analyzing blood flow inside the vena cava during these maneuvers. While during normal breathing, the vena cava flow can be considered almost stationary with a very low pressure gradient, during Valsalva the extravascular pressure compresses the vena cava resulting in a drastic reduction of the vein section, a global flow decrease through the cava but increasing the velocity magnitude. This change in the section is altered by the presence of the filter which forces the section of the vena cava before the renal veins to keep open. The effect of the presence of the filter is investigated during these maneuvers showing changes in wall shear stress and velocity patterns.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

3D computational parametric analysis of eccentric atheroma plaque: Influence of axial and circumferential residual stresses

  • Cilla, M.
  • Peña, E.
  • Martínez, M. A.
Plaque rupture plays a role in the majority of acute coronary syndromes. Rupture has usually been associated with stress concentrations, which are mainly affected by the plaque geometry and the tissue properties. The aim of this study is to evaluate the influence of morphology on the risk of plaque rupture, including the main geometrical factors, and to assess the role of circumferential and axial residual stresses by means of a parametric 3D finite element model. For this purpose, a 3D parametric finite element model of the coronary artery with eccentric atheroma plaque was developed. Healthy (adventitia and media in areas without atheroma plaque) and diseased (fibrotic and lipidic) tissues were considered in the model. The geometrical parameters used to define and design the idealized coronary plaque anatomy were the lipid core length, the stenosis ratio, the fibrous cap thickness, and the lipid core ratio. Finally, residual stresses in longitudinal and circumferential directions were incorporated into the model to analyse the influence of the important mechanical factors in the vulnerability of the plaque. Viewing the results, we conclude that residual stresses should be considered in the modelling of this kind of problems since they cause a significant alteration of the vulnerable plaque region limits. The obtained results show that the fibrous cap thickness and the lipid core length, in combination with the lipid core width, appear to be the key morphological parameters that play a determinant role in the maximal principal stress (MPS). However, the stenosis ratio is found to not play a significant role in vulnerability related to the MPS. Plaque rupture should therefore be observed as a consequence, not only of the cap thickness, but as a combination of the stenosis ratio, the fibrous cap thickness and the lipid core dimensions.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Does microcalcification increase the risk of rupture?

  • Cilla, M.
  • Monterde, D.
  • Pena, E.
  • Martinez, M. A.
Rupture of atherosclerotic plaque, which is related to maximal stress conditions in the plaque among others, is a major cause of mortality. More careful examination of stress distributions in atherosclerotic plaques reports that it could be due to local stress behaviors at critical sites caused by cap thinning, inflammation, macroscopic heterogeneity, and recently, the presence of microcalcifications. However, the role of microcalcifications is not yet fully understood, and most finite element models of blood vessels with atheroma plaque ignore the heterogeneity of the plaque constituents at the microscale. The goal of this work is to investigate the effect of microcalcifications on the stress field of an atheroma plaque vessel section. This is achieved by performing a parametric finite element study, assuming a plane strain hypothesis, of a coronary artery section with eccentric atheroma plaque and one microcalcification incorporated. The geometrical parameters used to define and design the idealized coronary plaque anatomy and the microcalcification were the fibrous cap thickness and the microcalcification ratio, angle and eccentricity. We could conclude that microcalcifications should be considered in the modeling of this kind of problems since they cause a significant alteration of the vulnerable risk by increasing the maximum maximal principal stress up to 32%, although this increase of stress is not uniform (12% on average). The obtained results show that the fibrous cap thickness, the microcalcification ratio and the microcalcification eccentricity, in combination with the microcalcification angle, appear to be the key morphological parameters that play a determinant role in the maximal principal stress and accordingly in the rupture risk of the plaque.