Skip to main content

DESARROLLO Y CALIBRACION DE UNA ETAPA PARA NANOPOSICIONADO 2D DE AMPLIO RANGO

DPI2010-21629-C02-01

•
Funding agency name Ministerio de Ciencia e Innovación
Funding agency acronym MICINN
Program Programa Nacional de Proyectos de Investigación Fundamental
Subprogram Subprograma de Proyectos de Investigación Fundamental no Orientada
Call Proyectos I+D
Call year 2010
State Plan 2008-2011 Plan Nacional de Investigación Científica, Desarrollo e Innovación Tecnológica
Management unit Subdirección General de Proyectos de Investigación
Beneficiary centre UNIVERSIDAD DE ZARAGOZA

Resultados relacionados

Resultados totales (Incluyendo duplicados): 5
Encontrada(s) 1 página(s)

Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Design optimization for the measurement accuracy improvement of a large range nanopositioning stage

  • Torralba Gracia, Marta
  • Yagüe Fabra, José Antonio
  • Albajez García, José Antonio
  • Aguilar Martín, Juan José
Both an accurate machine design and an adequate metrology loop definition are critical factors when precision positioning represents a key issue for the final system performance. This article discusses the error budget methodology as an advantageous technique to improve the measurement accuracy of a 2D-long range stage during its design phase. The nanopositioning platform NanoPla is here presented. Its specifications, e.g., XY-travel range of 50 mm ˆ 50 mm and sub-micrometric accuracy; and some novel designed solutions, e.g., a three-layer and two-stage architecture are described. Once defined the prototype, an error analysis is performed to propose improvement design features. Then, the metrology loop of the system is mathematically modelled to define the propagation of the different sources. Several simplifications and design hypothesis are justified and validated, including the assumption of rigid body behavior, which is demonstrated after a finite element analysis verification. The different error sources and their estimated contributions are enumerated in order to conclude with the final error values obtained from the error budget. The measurement deviations obtained demonstrate the important influence of the working environmental conditions, the flatness error of the plane mirror reflectors and the accurate manufacture and assembly of the components forming the metrological loop. Thus, a temperature control of ¿0.1 ¿C results in an acceptable maximum positioning error for the developed NanoPla stage, i.e., 41 nm, 36 nm and 48 nm in X-, Y- and Z-axis, respectively.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Teaching brain-machine interfaces as an alternative paradigm to neuroprosthetics control

  • Iturrate, I.
  • Chavarriaga, R.
  • Montesano, L.
  • Minguez, J.
  • Millán, J.D.R.
Brain-machine interfaces (BMI) usually decode movement parameters from cortical activity to control neuroprostheses. This requires subjects to learn to modulate their brain activity to convey all necessary information, thus imposing natural limits on the complexity of tasks that can be performed. Here we demonstrate an alternative and complementary BMI paradigm that overcomes that limitation by decoding cognitive brain signals associated with monitoring processes relevant for achieving goals. In our approach the neuroprosthesis executes actions that the subject evaluates as erroneous or correct, and exploits the brain correlates of this assessment to learn suitable motor behaviours. Results show that, after a short user € s training period, this teaching BMI paradigm operated three different neuroprostheses and generalized across several targets. Our results further support that these error-related signals reflect a task-independent monitoring mechanism in the brain, making this teaching paradigm scalable. We anticipate this BMI approach to become a key component of any neuroprosthesis that mimics natural motor control as it enables continuous adaptation in the absence of explicit information about goals. Furthermore, our paradigm can seamlessly incorporate other cognitive signals and conventional neuroprosthetic approaches, invasive or non-invasive, to enlarge the range and complexity of tasks that can be accomplished.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Preliminary Modelling and Implementation of the 2D-control for a Nanopositioning Long Range Stage

  • Torralba Gracia, Marta
  • Albajez García, José Antonio
  • Yagüe Fabra, José Antonio
  • Aguilar Martín, Juan José
Different systems are recently developed to obtain effective positioning at nanometer scale with increased working ranges. For this purpose, a two-dimensional nanopositioning platform (NanoPla) has been design and manufactured. To assure the demanding metrological performance the drive and control system is being defined and validated. Based on four home-made linear motors as actuators, this work is focused on the study of the control-loop for 1D- and 2D-cases with the aim of the preliminary modelling and posterior implementation. The different required blocks are presented and an initial controller solution is proposed to achieve the established positioning requests.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Geometrical characterisation of a 2D laser system and calibration of a cross-grid encoder by means of a self-calibration methodology

  • Torralba, M.
  • Díaz-Pérez, L.C.
  • Valenzuela, M.
  • Albajez, J.A.
  • Yagüe-Fabra, J.A.
This article presents a self-calibration procedure and the experimental results for the geometrical characterisation of a 2D laser system operating along a large working range (50 mm × 50 mm) with submicrometre uncertainty. Its purpose is to correct the geometric errors of the 2D laser system setup generated when positioning the two laser heads and the plane mirrors used as reflectors. The non-calibrated artefact used in this procedure is a commercial grid encoder that is also a measuring instrument. Therefore, the self-calibration procedure also allows the determination of the geometrical errors of the grid encoder, including its squareness error. The precision of the proposed algorithm is tested using virtual data. Actual measurements are subsequently registered, and the algorithm is applied. Once the laser system is characterised, the error of the grid encoder is calculated along the working range, resulting in an expanded submicrometre calibration uncertainty (k = 2) for the X and Y axes. The results of the grid encoder calibration are comparable to the errors provided by the calibration certificate for its main central axes. It is, therefore, possible to confirm the suitability of the self-calibration methodology proposed in this article.
Zaguán. Repositorio Digital de la Universidad de Zaragoza
Publication Artículo científico (JournalArticle)

Large range nanopositioning stage design: A three-layer and two-stage platform

  • Torralba, M.
  • Valenzuela , M.
  • Yagüe-Fabra, J. A.
  • Albajez, J. A.
  • Aguilar, J. J.
In this article, a novel two-dimensional nanopositioning platform (NanoPla) design is described. Its requirements are not only sub-micrometer accuracy for nanotechnology applications, but also long working range (XY-motion 50 mm × 50 mm). These features increase the common range operation of devices for nanotechnology issues (e.g. an atomic force microscope), and the number of potential metrological applications: positioning for manufacturing, manipulation or sample characterization. This novel design is characterized by a three-layer architecture and a two-stage motion strategy, which minimizes the measurement error. The manufactured prototype is here justified considering precision engineering principles and a wide state-of-art study of the literature, regarding long range nanopositioning stages. The simulations, the experimental results and the error budget also allowed, first, the optimization and, secondly, the validation of the design at nanometer scale.