MONITORIZACION Y DOSIMETRIA EN TERAPIA HADRONICA

PID2019-110657RB-I00

•
Funding agency name Agencia Estatal de Investigación
Funding agency acronym AEI
Program Programa Estatal de I+D+i Orientado a los Retos de la Sociedad
Call Proyectos de I+D+i (Generación de Conocimiento y Retos Investigación)
Call year 2019
State Plan 2017-2020 Plan Estatal de Investigación Científica y Técnica y de Innovación
Beneficiary centre AGENCIA ESTATAL CONSEJO SUPERIOR DE INVESTIGACIONES CIENTIFICAS (CSIC)

Resultados relacionados

Resultados totales (Incluyendo duplicados): 9
Encontrada(s) 1 página(s)

DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Contribución de congreso (ConferenceOutput)

Experimental characterization of MACACO III Compton telescope for hadron therapy treatment monitoring, Caracterización del telescopio Compton MACACO III para monitorización de terapia hadrónica

  • Barrientos, Luis
  • Borja-Lloret, Marina
  • Casaña, José Vicente
  • García López, J.
  • Hueso-González, F.
  • Jiménez-Ramos, M. C.
  • Lacasta Llácer, Carlos
  • Muñoz, Enrique
  • Ros García, Ana
  • Roser, J.
  • Senra, C.
  • Solaz, Carles
  • Viegas, R.
Trabajo presentado a las Jornadas de Física Médica, organizadas por la Real Sociedad Española de Física y el IFIC a través de la instalación de Física Médica IFIMED, celebradas virtualmente del 14 al 15 de diciembre de 2020., La terapia hadrónica es una técnica de tratamiento del cáncer consistente en la irradiación de tumores con protones u otros iones ligeros. El uso de esta técnica se ha incrementado por su ventaja de maximización de la dosis depositada sobre un tumor y de reducción de toxicidad del tratamiento comparado con la radioterapia convencional. Sin embargo, una de sus desventajas es la dificultad en la monitorización del rango del haz. Las cámaras Compton han resurgido como posibles candidatas para la monitorización de la terapia hadrónica, mediante la detección de los fotones producidos en la desexcitación de núcleos de los tejidos irradiados.El grupo IRIS del IFIC en Valencia se encuentra desarrollando un prototipo de una cámara Compton con la finalidad de ser empleada en monitorización de la terapia hadrónica. El prototipo MACACO (Medical Applications CompAct COmpton camera) se compone de cristales centelleadores de LaBr3 acoplados a una matriz de fotomultiplicadores de silicio. Con la finalidad de mejorar las prestaciones del prototipo, se ha construido una tercera versión que presenta un rendimiento mejorado respecto de sus predecesores. Los nuevos detectores de MACACO III se componen de cristales continuos de LaBr3 de 25,8 x 25,8 mm2 y 5 mm. El sistema de lectura AliVATA, desarrollado por el grupo, opera el ASIC VATA64HDR16 utilizado como electrónica de lectura. La caracterización de los detectores en el laboratorio se ha realizado utilizando fuentes radiactivas de 22Na, 152Eu y 137Cs. El mejor valor de resolución energética obtenido ha sido de 4.8 % FWHM a 662 keV, resolución angular de 6.0° FWHM a 1275 keV y eficiencia de detección en coincidencias de 1 x 10−3 a 1275 keV. A partir de los datos experimentales se han reconstruido imágenes de una fuente puntual de 22Na seleccionando eventos dobles y triples. El código de reconstrucción de MACACO III produce imágenes cuatridimensionales (energía y posición) usando eventos dobles o triples. Imágenes con la combinación de estos dos tipos de eventos ha sido recientemente conseguida. La FWHM de las imágenes reconstruidas fue de 2.4 mm y 3.3 mm FWHM a 1275 keV respectivamente a una distancia de la fuente de 70 mm. Con la finalidad de reproducir los resultados experimentales, MACACO III ha sido simulado en GATE v8.2 utilizando el módulo CCMod, obteniendo una buena correlación con los datos experimentales. MACACO III también ha sido caracterizado en aceleradores, con el fin de evaluar su respuesta en escenarios más realistas. Recientemente se ha testeado en el CNA (Sevilla) en un haz de protones de 18 MeV sobre un blanco de grafito. Los datos experimentales fueron tomados con el blanco en distintas posiciones y diferentes intensidades de haz. Imágenes con dos y tres planos fueron reconstruidas exitosamente en cada caso pudiendo distinguir distancias de 1 mm de separación del blanco. Futuras mejoras y pruebas en centros clínicos están en curso para alcanzar las prestaciones requeridas para la aplicación., This work has received funding from the Spanish Ministerio de Ciencia e Innovación (PID2019-110657RB-I00, FPA2017-85611-R and SEV-2014-0398), from Generalitat Valenciana (AICO/2019/070), from CSIC PIC 2018 programme (N2018FR0032) and from European Commission H2020 ENSAR2-MediNet (project number 654002). Group members are supported by Ramón y Cajal, SEJI and GENT programmes, UVEG Atracció de Talent, Generalitat Valenciana and Spanish Ministerio de Universidades FPU predoctoral contracts, Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Characterization of a Compton camera based on the TOFPET2 ASIC

  • Viegas, R.
  • Roser, J.
  • Barrientos, Luis
  • Borja-Lloret, Marina
  • Casaña, José Vicente
  • Jiménez-Ramos, M. C.
  • Hueso-González, F.
  • Ros García, Ana
  • Llosá, Gabriela
The use of Compton cameras for medical imaging and its interest as a hadron therapy treatment monitoring has increased in the last decade with the development of silicon photomultipliers. MACACO is a Compton camera prototype designed and assembled at the IRIS group of IFIC-Valencia. This Compton camera is based on monolithic Lanthanum (III) Bromide crystals and silicon photomultipliers, and employs the novel TOFPET2 ASIC as readout electronics. This system emerged as an alternative to MACACO II prototype, with the aim of improving its limited time resolution. To test the performance of the ASIC in a Compton camera setup, the prototype was characterized, both in laboratory and in-beam. A time resolution of 1.5 ns was obtained after time corrections, which improves greatly the performance of the MACACO II. Moreover, the results obtained at high photon energies demonstrate the ability of the system to obtain 1 mm displacements of the reconstructed spots. The results reinforce the potential of the system as a monitoring device for hadron therapy., This work has received funding from the Generalitat Valenciana, Spain (AICO/2019070) and from the Spanish Ministerio de Ciencia e Innovación (PID2019-110657RB-100). Group members are supported by UVEG Atracció de Talent, FPU, CDEIGENT, SEJIGENT and Generalitat Valenciana contracts. M.C. Jiménez-Ramos acknowledges the support to this work through a VI PPIT-US contract., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Hybrid PET/Compton-camera imaging: an imager for the next generation

  • Llosá, Gabriela
  • Rafecas López, Magdalena
Compton cameras can offer advantages over gamma cameras for some applications, since they are well suited for multitracer imaging and for imaging high-energy radiotracers, such as those employed in radionuclide therapy. While in conventional clinical settings state-of-the-art Compton cameras cannot compete with well-established methods such as PET and SPECT, there are specific scenarios in which they can constitute an advantageous alternative. The combination of PET and Compton imaging can benefit from the improved resolution and sensitivity of current PET technology and, at the same time, overcome PET limitations in the use of multiple radiotracers. Such a system can provide simultaneous assessment of different radiotracers under identical conditions and reduce errors associated with physical factors that can change between acquisitions. Advances are being made both in instrumentation developments combining PET and Compton cameras for multimodal or three-gamma imaging systems, and in image reconstruction, addressing the challenges imposed by the combination of the two modalities or the new techniques. This review article summarizes the advances made in Compton cameras for medical imaging and their combination with PET.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

System characterization and performance studies with MACACO III Compton camera

  • Barrientos, Luis
  • Borja-Lloret, Marina
  • Casaña, José Vicente
  • Hueso-González, F.
  • Ros, A.
  • Roser, J.
  • Senra, C.
  • Solaz, Carles
  • Viegas, R.
  • Llosá, Gabriela
The IRIS group of IFIC-Valencia has developed a Compton camera prototype. The system detectors are made of Lanthanum (III) bromide scintillator crystals coupled to silicon photomultipliers. Two models of silicon photomultipliers arrays with different micro pixel pitch (25 and 50 μm) have been chosen as possible candidates to improve the response of the new system. Characterization studies with a Na point-like source have indicated that the 25 μm photodetector provided better performance in terms of energy resolution (5.2% FWHM at 511 keV) and angular resolution (6.9 FWHM at 1275 keV), and more stability with temperature variations. In addition, MACACO III imaging capabilities have been assessed using a structure composed of thirty-seven Na point-like sources. Furthermore, in order to evaluate possible ways of improving the system performance, several studies have been carried out by means of simulations both in realistic and performance improved conditions. In this work, the system performance is evaluated for its future application in different areas.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Proton range verification with MACACO II Compton camera enhanced by a neural network for event selection

  • Muñoz, Enrique
  • Ros, A.
  • Borja-Lloret, Marina
  • Barrio, John
  • Dendooven, P.
  • Oliver, Josep F.
  • Ozoemelam, I.
  • Roser, J.
  • Llosá, Gabriela
The applicability extent of hadron therapy for tumor treatment is currently limited by the lack of reliable online monitoring techniques. An active topic of investigation is the research of monitoring systems based on the detection of secondary radiation produced during treatment. MACACO, a multi-layer Compton camera based on LaBr scintillator crystals and SiPMs, is being developed at IFIC-Valencia for this purpose. This work reports the results obtained from measurements of a 150 MeV proton beam impinging on a PMMA target. A neural network trained on Monte Carlo simulations is used for event selection, increasing the signal to background ratio before image reconstruction. Images of the measured prompt gamma distributions are reconstructed by means of a spectral reconstruction code, through which the 4.439 MeV spectral line is resolved. Images of the emission distribution at this energy are reconstructed, allowing calculation of the distal fall-off and identification of target displacements of 3 mm.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Gamma-ray sources imaging and test-beam results with MACACO III Compton camera

  • Barrientos, Luis
  • Borja-Lloret, Marina
  • Casaña, José Vicente
  • Dendooven, P.
  • García López, J.
  • Hueso-González, F.
  • Jiménez-Ramos, M. C.
  • Pérez-Curbelo, J.
  • Ros García, Ana
  • Roser, J.
  • Senra, C.
  • Viegas, R.
  • Llosá, Gabriela
© 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/)., Hadron therapy is a radiotherapy modality which offers a precise energy deposition to the tumors and a dose reduction to healthy tissue as compared to conventional methods. However, methods for real-time monitoring are required to ensure that the radiation dose is deposited on the target. The IRIS group of IFIC-Valencia developed a Compton camera prototype for this purpose, intending to image the Prompt Gammas emitted by the tissue during irradiation. The system detectors are composed of Lanthanum (III) bromide scintillator crystals coupled to silicon photomultipliers. After an initial characterization in the laboratory, in order to assess the system capabilities for future experiments in proton therapy centers, different tests were carried out in two facilities: PARTREC (Groningen, The Netherlands) and the CNA cyclotron (Sevilla, Spain). Characterization studies performed at PARTREC indicated that the detectors linearity was improved with respect to the previous version and an energy resolution of 5.2 % FWHM at 511 keV was achieved. Moreover, the imaging capabilities of the system were evaluated with a line source of 68Ge and a point-like source of 241Am-9Be. Images at 4.439 MeV were obtained from irradiation of a graphite target with an 18 MeV proton beam at CNA, to perform a study of the system potential to detect shifts at different intensities. In this sense, the system was able to distinguish 1 mm variations in the target position at different beam current intensities for measurement times of 1800 and 600 s., This work has received funding from the Spanish Ministerio de Ciencia e Innovación/Agencia Estatal de Investigación (PID2019-110657RB-I00) and from the European Union’s Horizon 2020 research and innovation programme under grant agreement No 654002. Group members are supported by Generalitat Valenciana CDEIGENT programme, UVEG Atracció de Talent, Generalitat Valenciana, Spanish Ministerio de Universidades FPU, Becas Chile and Grisolía predoctoral contracts . M.C. Jiménez-Ramos acknowledges the support to this work through a VI PPIT-US contract., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Influence of the background in Compton camera images for proton therapy treatment monitoring

  • Borja-Lloret, Marina
  • Barrientos, Luis
  • Bernabéu, José
  • Lacasta Llácer, Carlos
  • Muñoz, Enrique
  • Ros García, Ana
  • Roser, J.
  • Viegas, R.
  • Llosá, Gabriela
Objective. Background events are one of the most relevant contributions to image degradation in Compton camera imaging for hadron therapy treatment monitoring. A study of the background and its contribution to image degradation is important to define future strategies to reduce the background in the system. Approach. In this simulation study, the percentage of different kinds of events and their contribution to the reconstructed image in a two-layer Compton camera have been evaluated. To this end, GATE v8.2 simulations of a proton beam impinging on a PMMA phantom have been carried out, for different proton beam energies and at different beam intensities. Main results. For a simulated Compton camera made of Lanthanum (III) Bromide monolithic crystals, coincidences caused by neutrons arriving from the phantom are the most common type of background produced by secondary radiations in the Compton camera, causing between 13% and 33% of the detected coincidences, depending on the beam energy. Results also show that random coincidences are a significant cause of image degradation at high beam intensities, and their influence in the reconstructed images is studied for values of the time coincidence windows from 500 ps to 100 ns. Significance. Results indicate the timing capabilities required to retrieve the fall-off position with good precision. Still, the noise observed in the image when no randoms are considered make us consider further background rejection methods.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

Improving Compton camera imaging of multi-energy radioactive sources by using machine learning algorithms for event selection

  • Pérez-Curbelo, J.
  • Roser, J.
  • Muñoz, Enrique
  • Barrientos, Luis
  • Sanz, Verónica
  • Llosá, Gabriela
Event selection and background reduction for Compton camera imaging of multi-energy radioactive sources has been performed by employing neural networks. A Compton camera prototype with detectors made of LaBr<inf>3</inf> crystals coupled to silicon photomultiplier arrays was used to acquire experimental data from a circular array of <sup>22</sup>Na sources. The prototype and two arrays of <sup>22</sup>Na sources were simulated with GATE v8.2 Monte Carlo code, to obtain data for neural network training. Neural network models were trained on simulated data for event classification. The optimum models were found by using Weights & Biases platform tools. The trained models were used to classify simulated and real data for selecting signal events and rejecting background prior to image reconstruction. The models performed well on simulated data. The image obtained with experimental data showed an improvement with respect to event selection with energy cuts. The method is promising for Compton camera imaging of multi-energy radioactive sources., This work has received funding from Ministerio de Ciencia e Innovación, Agencia Estatal de Investigación (PID2019-110657RB-I00) and from the Generalitat Valenciana, Spain (GRISOLIAP/2021/099)., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publication Artículo científico (JournalArticle)

System modeling and image reconstruction algorithm optimization for a Compton camera for hadron therapy range verification

  • Roser, J.
  • Llosá, Gabriela
Compton cameras are electronically collimated devices which offer a compact solution for multi-energy gamma-ray imaging. For this reason, such devices are being investigated as potential tools to achieve online range verification in hadron therapy, through the so-called Prompt Gamma imaging technique. However, this application is notably challenging, entailing the need of complex image reconstruction procedures and requiring excellent detector performance in order to achieve sufficient image quality. In this work we recapitulate the steps made in the recent years in system modeling and adaptation of existing image reconstruction algorithms towards the full exploitation of MACACO, a multi-layer Compton telescope under development at the IRIS group of IFIC (Valencia). © The Author(s) 2025., This work has received funding from the Spanish Ministerio de Ciencia, Innovación y Universidades, Agencia Estatal de Investigación (PID2019-110657RB-I00 and PID2022-143246OB-I00)., Peer reviewed