Publication Artículo científico (JournalArticle)

Interoperability in digital electrocardiography: Harmonization of ISO/IEEE x73-PHD and SCP-ECG

  • Trigo Vilaseca, Jesús Daniel
  • Chiarugi, Franco
  • Alesanco, Álvaro
  • Martínez de Espronceda Cámara, Miguel
  • Serrano Arriezu, Luis Javier
  • Chronaki, Catherine E
  • Escayola Calvo, Javier
  • Martínez Ruiz, Ignacio
  • García, José
The ISO/IEEE 11073 (x73) family of standards is a reference frame for medical device interoperability. A draft for an ECG device specialization (ISO/IEEE 11073-10406-d02) has already been presented to the Personal Health Device (PHD) Working Group, and the Standard Communications Protocol for Computer-Assisted ElectroCardioGraphy (SCP-ECG) Standard for short-term diagnostic ECGs (EN1064:2005+A1:2007) has recently been approved as part of the x73 family (ISO 11073-91064:2009). These factors suggest the coordinated use of these two standards in foreseeable telecardiology environments, and hence the need to harmonize them. Such harmonization is the subject of this paper. Thus, a mapping of the mandatory attributes defined in the second draft of the ISO/IEEE 11073-10406-d02 and the minimum SCP-ECG fields is presented, and various other capabilities of the SCP-ECG Standard (such as the messaging part) are also analyzed from an x73-PHD point of view. As a result, this paper addresses and analyzes the implications of some inconsistencies in the coordinated use of these two standards. Finally, a proof-of-concept implementation of the draft x73-PHD ECG device specialization is presented, along with the conversion from x73-PHD to SCP-ECG. This paper, therefore, provides recommendations for future implementations of telecardiology systems that are compliant with both x73-PHD and SCP-ECG., This work was supported in part by the Innovation and Science Ministry (MICINN) and the European Regional Development Fund under Project TIN-2011-23792, and in part by the Ministry of Industry, Tourism and Trade (MITyC) under Project TSI-020100-2010-277.
Academica-e. Repositorio Institucional de la Universidad Pública de Navarra
Publication Capítulo o parte de un libro (BookPart)

Legislation, standardization and technological solutions for enhancing e-Accessibility in e-Health

  • Valle García, Pilar del
  • Martínez Ruiz, Ignacio
  • Escayola Calvo, Javier
  • Trigo Vilaseca, Jesús Daniel
  • García Moros, José
This chapter presents a comprehensive review of the most recent advances in legislation, standardization and technological solutions for enhancing e-Accessibility in e-Health. In Section 2, the current state of the legislation in this field is reviewed. Section 3 describes and analyzes the main characteristics of ISO 9241, the standard that regulates the legal implementation of the design requirements for e-Accessibility. Finally, the recommended technology requirements, as well as the specialized medical devices and products for each type of disability, are presented and discussed in Section 4., This work was partially supported by projects TIN2008-00933/TSI of the Innovation and Science Ministry (MICINN) and European Funds for Regional Development (EFRD), TSI-020100-2010-277 and TSI-020302-2009-7/Plan Avanza I+D of Ministry of Industry, Tourism and Trade.
Academica-e. Repositorio Institucional de la Universidad Pública de Navarra
Publication Artículo científico (JournalArticle)

Implementation methodology for interoperable personal health devices with low-voltage low-power constraints

  • Martínez de Espronceda Cámara, Miguel
  • Serrano Arriezu, Luis Javier
  • Led Ramos, Santiago
  • Trigo Vilaseca, Jesús Daniel
  • Marzo Pérez, Asier
  • Escayola Calvo, Javier
  • García, José Ángel
Traditionally, e-Health solutions were located at the point of care (PoC), while the new ubiquitous user-centered paradigm draws on standard-based personal health devices (PHDs). Such devices place strict constraints on computation and battery efficiency that encouraged the International Organization for Standardization/IEEE11073 (X73) standard for medical devices to evolve from X73PoC to X73PHD. In this context, low-voltage low-power (LV-LP) technologies meet the restrictions of X73PHD-compliant devices. Since X73PHD does not approach the software architecture, the accomplishment of an efficient design falls directly on the software developer. Therefore, computational and battery performance of such LV-LP-constrained devices can even be outperformed through an efficient X73PHD implementation design. In this context, this paper proposes a new methodology to implement X73PHD into microcontroller-based platforms with LV-LP constraints. Such implementation methodology has been developed through a patterns-based approach and applied to a number of X73PHD-compliant agents (including weighing scale, blood pressure monitor, and thermometer specializations) and microprocessor architectures (8, 16, and 32 bits) as a proof of concept. As a reference, the results obtained in the weighing scale guarantee all features of X73PHD running over a microcontroller architecture based on ARM7TDMI requiring only 168 B of RAM and 2546 B of flash memory., This work was supported in part by Comisión Interministerial de Ciencia y Tecnología (CICYT) and European Regional Development Fund (ERDF) under Project TIN-2009-08414 and Project TIN-2008-00933/TSI, and by Ministerio de Industria, Turismo y Comercio, and FPI Grant to M. Martínez-Espronceda (Res.1342/2006 Public University of Navarre) under Project TSI-020100-2010-277, Project TSI-020302-2009-89, and Project TSI-020302-2009-7/Plan Avanza I+D.