Feasibility Study of a Small Animal PET Insert Based on a Single LYSO Monolithic Tube

  • González Martínez, Antonio Javier
  • F Sánchez
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
  • Berr, Stuart S.
  • Cañizares-Ledo, Gabriel
  • Gonzalez-Montoro, Andrea
  • Orero Palomares, Abel
  • Correcher Salvador, Carlos
  • Rezaei, Ahmadreza
  • Nuyts, Johan
  • Majewski, Stan
[EN] There are drawbacks with using a Positron Emission Tomography (PET) scanner design employing the traditional arrangement of multiple detectors in an array format. Typically PET systems are constructed with many regular gaps between the detector modules in a ring or box configuration, with additional axial gaps between the rings. Although this has been significantly reduced with the use of the compact high granularity SiPM photodetector technology, such a scanner design leads to a decrease in the number of annihilation photons that are detected causing lower scanner sensitivity. Moreover, the ability to precisely determine the line of response (LOR) along which the positron annihilated is diminished closer to the detector edges because the spatial resolution there is degraded due to edge effects. This happens for both monolithic based designs, caused by the truncation of the scintillation light distribution, but also for detector blocks that use crystal arrays with a number of elements that are larger than the number of photosensors and, therefore, make use of the light sharing principle. In this report we present a design for a small-animal PET scanner based on a single monolithic annulus-like scintillator that can be used as a PET insert in high-field Magnetic Resonance systems. We provide real data showing the performance improvement when edge-less modules are used. We also describe the specific proposed design for a rodent scanner that employs facetted outside faces in a single LYSO tube. In a further step, in order to support and prove the proposed edgeless geometry, simulations of that scanner have been performed and lately reconstructed showing the advantages of the design., This project was funded in part by the European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program (grant agreement No 695536). It has also been supported by the Spanish Ministerio de Economia, Industria y Competitividad under Grant TEC2016-79884-C2-1-R and through PROSPET (DTS15/00152) funded by the Ministerio de Economia y Competitividad. AR is a postdoctoral fellow of the FWO (project 12T7118N). The University of Virginia School of Medicine has provided seed funding for this project.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

2D feasibility study of joint reconstruction of attenuation and activity in limited angle TOF-PET

  • Vergara, Marina
  • Rezaei, Ahmadreza
  • Schramm, Georg
  • Nuyts, Johan
  • Rodríguez-Álvarez, M.J.|||0000-0001-8333-8792
  • Benlloch Baviera, Jose María|||0000-0001-6073-1436
[EN] Several research groups are studying organ-dedicated limited angle positron emission tomography (PET) systems to optimize performance-cost ratio, sensitivity, access to the patient, and/or flexibility. Often open systems are considered, typically consisting of two detector panels of various sizes. Such systems provide incomplete sampling due to limited angular coverage and/or truncation, which leads to artifacts in the reconstructed activity images. In addition, these organ-dedicated PET systems are usually stand-alone systems, and as a result, no attenuation information can be obtained from anatomical images acquired in the same imaging session. It has been shown that the use of time-of-flight (TOF) information reduces incomplete data artifacts and enables the joint estimation of the activity and the attenuation factors. In this work, we explore with simple 2-D simulations the performance and stability of a joint reconstruction algorithm, for imaging with a limited angle PET system. The reconstruction is based on the so-called maximum-likelihood attenuation correction factors (MLACF) algorithm and uses linear attenuation coefficients in a known-tissue-class region to obtain absolute quantification. Different panel sizes and different TOF resolutions are considered. The noise propagation is compared to that of MLEM reconstruction with exact attenuation correction (AC) for the same PET system. The results show that with good TOF resolution, images of good visual quality can be obtained. If also a good scatter correction can be implemented, quantitative PET imaging will be possible. Further research, in particular on scatter correction, is required., This work was supported in part by the European Research Council (ERC) through the European Unions Horizon 2020 Research and Innovation Program under Grant 695536; in part by the Research Foundation Flanders (FWO) under Project 12T7118N and Project G062220N; and in part by the NIH Project under Grant 1P41EB017183-01A1.