Publicación Artículo científico (JournalArticle)

Hybrid compton-PET imaging for ion-range verification: a preclinical study for proton, helium, and carbon therapy at HIT

  • Balibrea-Correa, Javier
  • Lerendegui-Marco, Jorge
  • Ladarescu, Ion
  • Guerrero, Carlos
  • Rodríguez-González, Teresa
  • Jiménez-Ramos, Maria del Carmen
  • Quesada, Jose Manuel
  • Bauer, Julia
  • Brons, Stephan
  • Domingo-Pardo, César
  • Morell-Ortega, Sergio|||0009-0001-7639-2915
[EN] Enhanced-accuracy ion-range verification in real time shall enable a significant step forward in the use of therapeutic ion beams. Positron-emission tomography (PET) and prompt-gamma imaging (PGI) are two of the most promising and researched methodologies, both of them with their own advantages and challenges. Thus far, both of them have been explored for ion-range verification in an independent way. However, the simultaneous combination of PET and PGI within the same imaging framework may open-up the possibility to exploit more efficiently all radiative emissions excited in the tissue by the ion beam. Here, we report on the first preclinical implementation of an hybrid PET-PGI imaging system, hereby exploring its performance over several ion beam species (H, He and C), energies (55¿275 MeV) and intensities ( - ions/spot), which are representative of clinical conditions. The measurements were carried out using the pencil-beam scanning technique at the synchrotron accelerator of the heavy ion therapy center in Heidelberg utilizing an array of four Compton cameras in a twofold front-to-front configuration. The results demonstrate that the hybrid PET-PGI technique can be well suited for relatively low energies (55¿155 MeV) and beams of protons. On the other hand, for heavier beams of helium and carbon ions at higher energies (155¿275 MeV), range monitoring becomes more challenging owing to large backgrounds from additional nuclear processes. The experimental results are well understood on the basis of realistic Monte Carlo calculations, which show a satisfactory agreement with the measured data. This work can guide further upgrades of the hybrid PET-PGI system toward a clinical implementation of this innovative technique., This work derives from fundamental research carried out with funding from the European Research Council (ERC) under the European Union s Horizon 2020 research and innovation program (ERC Consolidator Grant project HYMNS, with grant agreement No. 681740) and part of this work has been funded under ERC Proof-of-Concept Grant GNVISION (Grant Agreement 101113330). The authors acknowledge support from the Spanish Ministerio de Ciencia e Innovación under grants PDC2021-121536-C21, PID2019-104714GB-C21 and RTI2018-098117-B-C21. We acknowledge support from the Severo Ochoa Grant CEX2023-001292-S funded by MCIU/AEI. T. Rodriguez-Gonzalez acknowledges the Spanish FPI predoctoral grant. J. Balibrea-Correa is supported by grant ICJ220-045122-I and J. Lerendegui-Marco is supported by grant FJC2020-044688-I funded by MCIN/AEI/10.13039/501100011033 and CIAPOS/2022/020 funded by MICIN/AEI and Generalitat Valenciana. Ion Ladarescu was funded by the R&D grant PDC2022-133382-I00, funded by the Spanish Ministry of Science (MCIN/AEI/10.13039/501100011033). M.C. Jiménez-Ramos acknowledges the support to this work through a VI PPIT-US contract. We would like to thank the crew at the Electronics Laboratory of IFIC, in particular Manuel Lopez Redondo and Jorge Nácher Arándiga for their excellent and efficient work prototyping and manufacturing customized PCBs for this work.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publicación Artículo científico (JournalArticle)

Hybrid in-beam PET- and Compton prompt-gamma imaging aimed at enhanced proton-range verification

  • Balibrea, Javier
  • Lerendegui-Marco, J.
  • Ladarescu, I.
  • Guerrero, C.
  • Rodríguez González, M. Teresa
  • Jiménez-Ramos, M. C.
  • Fernández-Martínez, Begoña
  • Domingo Pardo, César
We report on a hybrid in-beam PET and prompt-gamma Compton imaging system aimed at quasi real-time ion-range verification in proton-therapy treatments. Proof-of-concept experiments were carried out at the radiobiology beam line of the CNA cyclotron facility using a set of two synchronous Compton imagers and different target materials. The time structure of the 18 MeV proton beam was shaped with a series of beam-on and beam-off intervals, thereby mimicking a pulsed proton beam on a long time scale. During beam-on intervals, Compton imaging was performed utilizing the high energy γ-rays promptly emitted from the nuclear reactions occurring in the targets. In the course of the beam-off intervals in situ positron-emission tomography was accomplished with the same imagers using the β decay of activated nuclei. The targets used were stacks of different materials covering also various proton ranges and energies. A systematic study on the performance of these two complementary imaging techniques is reported and the experimental results interpreted on the basis of Monte Carlo calculations. The results demonstrate the possibility to combine both imaging techniques in a concomitant way, where high-efficiency prompt-gamma imaging is complemented with the high spatial accuracy of PET. Empowered by these results we suggest that a pulsed beam with a suitable duty cycle, in conjunction with in situ Compton- and PET-imaging may help to attain complementary information and quasi real-time range monitoring with high accuracy., Open Access funding provided thanks to the CRUE-CSIC agreement with springer nature. This work has been carried out in the framework of a project funded by the European Research Council (ERC) under the European Union’s Horizon 2020 research and innovation programme (ERC Consolidator Grant project HYMNS, with grant agreement No.~681740). The authors acknowledge support from the Spanish Ministerio de Ciencia e Innovación under grants PID2019-104714GB-C21, FPA2017-83946-C2-1-P, FIS2015-71688-ERC, FPA2016-77689-C2-1-R, RTI2018-098117-B-C21, CSIC for funding PIE-201750I26 and from the European H2020-847552 (SANDA). This work was partially supported by Generalitat Valenciana PROMETEO/2019/007. J. Balibrea-Correa is supported by grant ICJ220-045122-I funded by MCIN/AEI/10.13039/501100011033 and by European Union NextGenerationEU/PRTR. J. Lerendegui-Marco is supported by grant FJC2020-044688-I funded by MCIN/AEI/10.13039/501100011033 and by European UnionNextGenerationEU/PRTR.