High performance detector head for PET and PET/MR with continuous crystals and SiPMs

  • Llosá, G.
  • Barrillon, P.
  • Bisogni, M. G.
  • Cabello, J.
  • Del Guerra, A.
  • Etxebeste, A.
  • Gillam, J.E.
  • Lacasta, C.
  • Rafecas, M.
  • Solaz, C.
  • Stankova, V.
  • de la Taille, C.
  • Barrio-Toala, John|||0000-0001-5181-2146
  • Oliver-Guillén, José Francisco|||0000-0002-8585-9345
[EN] A high resolution PET detector head for small animal PET applications has been developed. The detector is composed of a 12 mm x 12 mm continuous LYSO crystal coupled to a 64-channel monolithic SiPM matrix from FBK-irst. Crystal thicknesses of 5 mm and 10 mm have been tested, both yielding an intrinsic spatial resolution around 0.7 mm FWHM with a position determination algorithm that can also provide depth-of-interaction information. The detectors have been tested in a rotating system that makes it possible to acquire tomographic data and reconstruct images of Na-22 sources. An image reconstruction method specifically adapted for continuous crystals has been employed. The Full Width at Half Maximum measured from a point source reconstructed with ML-EM was 0.7 mm with the 5 mm crystal and 0.8 mm with the 10 mm crystal., This work is partially supported by the European Commission's 7th Framework Programme through a Marie Curie European Reintegration Grant (ASPID GA NUM 239362), and by the Spanish Ministerio de Economia y Competitividad/Plan Nacional de I + D + i Grants TEC2007-61047 and FPA2010-14891. Several group members are supported through the Juan de la Cierva, JAE-DOC and Atraccio de Talent (UV) programs.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

High resolution detectors based on continuous crystals and SiPMs for small animal PET

  • Cabello, J.
  • Barrillon, P.
  • Bisogni, M. G.
  • Del Guerra, A.
  • Lacasta, C.
  • Rafecas, M.
  • Saikouk, H.
  • Solaz, C.
  • Solevi, P.
  • de la Taille, C.
  • Llosá, G.
  • Barrio-Toala, John|||0000-0001-5181-2146
[EN] Sensitivity and spatial resolution are the two main factors to maximize in emission imaging. The improvement of one factor deteriorates the other with pixelated crystals. In this work we combine SiPM matrices with monolithic crystals, using an accurate g-ray interaction position determination algorithm that provides depth of interaction. Continuous crystals provide higher sensitivity than pixelated crystals, while an accurate interaction position determination does not degrade the spatial resolution. Monte Carlo simulations and experimental data show good agreement both demonstrating submillimetre intrinsic spatial resolution. A system consisting in two rotating detectors in coincidence is currently under operation already producing tomographic images., This work is partially supported by the European Commission s 7th Framework Programme through a Marie Curie European Reintegration Grant (ASPID GA NUM 239362), and by the Spanish Ministerio de Ciencia e Innovacio¿n/Plan Nacional de IþDþi Grants TEC2007-61047 and FPA2010-14891. G. Llosa¿ was partly funded by a JAE DOC Grant. H. Saikouk was supported by the Spanish PCI2011 A/035250/11 Project.
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Publicación Artículo científico (JournalArticle)

Application of Artificial Neural Network for Reducing Random Coincidences in PET

  • Oliver, J,F.
  • Cabello, J.
  • Tortajada Velert, Salvador
  • Rafecas Lopez, Maria Magdalena
  • Fuster García, Elíes|||0000-0002-0716-8960
Positron Emission Tomography (PET) is based on the detection in coincidence of the two photons created in a positron annihilation. In conventional PET, this coincidence identification is usually carried out through a coincidence electronic unit. An accidental coincidence occurs when two photons arising from different annihilations are classified as a coincidence. Accidental coincidences are one of the main sources of image degradation in PET. Some novel systems allow coincidences to be selected post-acquisition in software, or in real time through a digital coincidence engine in an FPGA. These approaches provide the user with extra flexibility in the sorting process and allow the application of alternative coincidence sorting procedures. In this work a novel sorting procedure based on Artificial Neural Network (ANN) techniques has been developed. It has been compared to a conventional coincidence sorting algorithm based on a time coincidence window. The data have been obtained from Monte-Carlo simulations. A small animal PET scanner has been implemented to this end. The efficiency (the ratio of correct identifications) can be selected for both methods. In one case by changing the actual value of the coincidence window used, and in the other by changing a threshold at the output of the neural network. At matched efficiencies, the ANN-based method always produces a sorted output with a smaller random fraction. In addition, two differential trends are found: the conventional method presents a maximum achievable efficiency, while the ANN-based method is able to increase the efficiency up to unity, the ideal value, at the cost of increasing the random fraction. Images reconstructed using ANN sorted data (no compensation for randoms) present better contrast, and those image features which are more affected by randoms are enhanced. For the image quality phantom used in the paper, the ANN method decreases the spillover ratio by a factor of 18%., This work was supported by RD12/0036/0020 Red Tematica de Investigacion Coperativa en Cancer (RTICC) 2012; Ministerio de Ciencia e Innovacion through a JAE-Doc contract associated to CSIC and FSE co-financed; also under projects TEC-2007-61047 and FPA-2010-14891.
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publicación Artículo científico (JournalArticle)

Improving the singles rate method for modeling accidental coincidences in high-resolution PET

  • Oliver, Josep F.
  • Rafecas, Magdalena
Random coincidences ('randoms') are one of the main sources of image degradation in PET imaging. In order to correct for this effect, an accurate method to estimate the contribution of random events is necessary. This aspect becomes especially relevant for high-resolution PET scanners where the highest image quality is sought and accurate quantitative analysis is undertaken. One common approach to estimate randoms is the so-called singles rate method (SR) widely used because of its good statistical properties. SR is based on the measurement of the singles rate in each detector element. However, recent studies suggest that SR systematically overestimates the correct random rate. This overestimation can be particularly marked for low energy thresholds, below 250 keV used in some applications and could entail a significant image degradation. In this work, we investigate the performance of SR as a function of the activity, geometry of the source and energy acceptance window used. We also investigate the performance of an alternative method, which we call 'singles trues' (ST) that improves SR by properly modeling the presence of true coincidences in the sample. Nevertheless, in any real data acquisition the knowledge of which singles are members of a true coincidence is lost. Therefore, we propose an iterative method, STi, that provides an estimation based on ST but which only requires the knowledge of measurable quantities: prompts and singles. Due to inter-crystal scatter, for wide energy windows ST only partially corrects SR overestimations. While SR deviations are in the range 86-300% (depending on the source geometry), the ST deviations are systematically smaller and contained in the range 4-60%. STi fails to reproduce the ST results, although for not too high activities the deviation with respect to ST is only a few percent. For conventional energy windows, i.e. those without inter-crystal scatter, the ST method corrects the SR overestimations, and deviations from the true random rate are of the order of 1% or less. In addition, in the case of conventional energy window STi results reproduce ST results and therefore the former can be used to obtain the true random rate., This work was supported by the Spanish Ministry of Science and Innovation under grants PCI2006-A7-0602 and TEC2007-61047. One of the authors, JFO, has been supported by the Generalitat Valenciana through the contract APOSTD/2007/026 and by the Consejo Superior de Investigaciones Científicas (CSIC) through a contract in the JAE-Doc modality of the program ‘Junta para la Ampliación de Estudios’ co-financed by FSE., Peer reviewed
DIGITAL.CSIC. Repositorio Institucional del CSIC
Publicación Artículo científico (JournalArticle)

The AX-PET demonstrator - Design, construction and characterization

  • Beltrame, P.
  • Bolle, E.
  • Braem, A.
  • Casella, C.
  • Chesi, E.
  • Clinthorne, N.
  • De Leo, R.
  • Dissertori, G.
  • Djambazov, L.
  • Fanti, V.
  • Heller, M.
  • Joram, C.
  • Kagan, H.
  • Lustermann, W.
  • Meddi, F.
  • Nappi, E.
  • Nessi-Tedaldi, F.
  • Oliver, Josep F.
  • Pauss, F.
  • Rafecas, Magdalena
  • Renker, D.
  • Rudge, A.
  • Schinzel, D.
  • Schneider, T.
  • Séguinot, J.
  • Solevi, P.
  • Stapnes, S.
  • Weilhammer, P.
Axial PET is a novel geometrical concept for Positron Emission Tomography (PET), based on layers of long scintillating crystals axially aligned with the bore axis. The axial coordinate is obtained from arrays of wavelength shifting (WLS) plastic strips placed orthogonally to the crystals. This article describes the design, construction and performance evaluation of a demonstrator set-up which consists of two identical detector modules, used in coincidence. Each module comprises 48 LYSO crystals of 100 mm length and 156 WLS strips. Crystals and strips are readout by Geiger-mode Avalanche Photo Diodes (G-APDs). The signals from the two modules are processed by fully analog front-end electronics and recorded in coincidence by a VME-based data acquisition system. Measurements with point-like <sup>22</sup>Na sources, with the modules used both individually and in coincidence mode, allowed for a complete performance evaluation up to the focal plane reconstruction of point sources. The results obtained are in good agreement with expectations and proved the set-up to be ready for the next evaluation phase with PET phantoms filled with radiotracers. © 2011 Elsevier B.V. All right reserved., The work is partially funded through the Spanish Ministry of Science & Innovation under Grant FPA2008-02419-E and project Grant TEC2007-61047. P. Solevi is supported by the Marie Curie Intra-European Fellowship Grant 237620 for the project INSPET. P. Beltrame is supported through the Marie Curie Initial Training Network MC-PAD (GA no. 214560-2)., No